Methamphetamine and Illicit Drugs, Precursors and Adulterants on Wipes by Liquid-Liquid Extraction (9106)

NIOSH Manual of Analytical Methods (1994)
National Institute for Occupational Safety and Health
Methamphetamine and Illicit Drugs, Precursors and Adulterants on Wipes by Liquid-Liquid Extraction (9106)
2003118NIOSH Manual of Analytical Methods — Methamphetamine and Illicit Drugs, Precursors and Adulterants on Wipes by Liquid-Liquid Extraction (9106)1994National Institute for Occupational Safety and Health

METHAMPHETAMINE and Illicit Drugs, Precursors and Adulterants on Wipes by Liquid-Liquid Extraction FORMULA: Table 1

MW: Table 1

METHOD: 9106, Issue 1

CAS: Table 1

EVALUATION: Partial

U.S. regulatory OELS OSHA or MSHA: none for surfaces Other published OELs and guidelines NIOSH, ACGIH, or AIHA: none for surfaces States: Table 2


9106

RTECS: Table 1

Issue 1: 17 October 2011

PROPERTIES: Table 3

SYNONYMS: Table 4 SAMPLING SAMPLER:

Wipe

SAMPLE AREA:

100 cm² or 1000 cm²

SHIPMENT:

Preferably ship refrigerated, <6 °C

SAMPLE STABILITY:

30 days at <6 °C (See Table 5)

FIELD BLANKS:

2 to 10 blanks per sample set

MEASUREMENT ACCURACY LEVEL STUDIED:

3 μg/sample on smooth surfaces

BIAS:

Table 11a and 11b

OVERALL PRECISION ( ACCURACY:

):

Table 8a and 8b [1] Table 8a and 8b [1]

MEASUREMENT TECHNIQUE:

GAS CHROMATOGRAPHY/MASS SPECTROMETRY

ANALYTES:

Table 1

DESORPTION:

0.1 M sulfuric acid

CLEANUP/ EXTRACTION:

Hexane cleanup followed by methylene chloride extraction

DERIVATIZATION:

Chlorodifluoroacetic anhydride

INJECTION:

2 µL, splitless

TEMPERATURE - Injection: 265 °C - Detector: 285 °C - Column: 90 °C (2 min), 310 °C (10 °C/ min), hold for 11 min MASS SPECTROMETER: Scan mode (29 – 470 AMU), 2 scan per sec, Selected ion monitoring (SIM) mode Table 6 CARRIER GAS:

Helium, 1.5 mL/min

COLUMN:

Capillary, fused silica, 30 m × 0.32 mm ID; 0.5 μm film DB-5ms or equivalent

CALIBRATION:

Standards from spiked wipes with internal standard, See Table 7

RANGE:

Table 8a and 8b [1]

ESTIMATED LOD:

Table 5

PRECISION ( ):

Table 8a and 8b [1]

APPLICABILITY: For methamphetamine, the range is 0.05 to 60 μg/sample (sample = 100 cm2 or 1000 cm2). This method was developed for the analysis of selected drugs and precursors on surfaces in clandestine drug labs. [1, 2] Sampling methodology was tested using wipes on smooth, non-porous surfaces. The APPENDIX contains sampling information for other types of surfaces. INTERFERENCES: No chromatographic interferences detected. Water, surfactants and polyols inhibit derivatization. OTHER WIPE METHODS: NIOSH 9109 uses solid-phase extraction and gas chromatography/mass spectrometry (GC/MS) to measure multiple drugs [3]. NIOSH 9111 uses liquid chromatography/mass spectrometry (LC/MS) to measure methamphetamine [4]. NIOSH Manual of Analytical Methods (NMAM), Fifth Edition

Method rev. 1.1.1 METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 2 of 31

REAGENTS: NOTE: See APPENDIX A for special instructions on reagents. 1. Analytes listed in Table 1.* 2. Internal standards from those listed in Table 9. 3. Solvents, residue free analytical grades: a. Hexane * b. Isopropanol (IPA)* c. Methanol * d. Methylene chloride (CH2Cl2) * e. Toluene * f. Acetone* 4. Concentrated sulfuric and hydrochloric acids (AR or trace metals analysis grades).* 5. Sodium hydroxide, A.C.S. grade.* 6. Anhydrous granular sodium sulfate, AR grade. 7. Anhydrous granular potassium carbonate, AR grade. 8. Bromothymol Blue, ≥95%, A.C.S., phenolphthalein, A.C.S.; crystal violet (Gentian Violet), ≥95%, A.C.S. 9. Purified gases: helium for carrier gas, nitrogen for drying. 10. Chlorodifluoroacetic anhydride, 98%* derivatizing agent. Moisture sensitive! 11. 4,4’-Dibromooctafluorobiphenyl, 99%, instrument internal standard (IIS). 12. Deionized water (ASTM type II).

SOLUTIONS: NOTE: See APPENDIX A for special instructions on solutions. 1. Prepare solutions of analytes of interest. Calculate concentrations as the free base. Keep refrigerated (<6 °C). Protect solutions from light. a. Stock solutions are prepared at about 1-2 mg/mL in methanol. b. Analyte spiking solutions are prepared by diluting the stock solutions to about 200 μg/mL each in methanol. 2. Prepare internal standard spiking solution in methanol at about 200 μg/mL. (Note: Add about 2 milligrams of crystal violet per 20 mL of internal standard spiking solution to help indicate which samples have been spiked.) 3. Desorption solution: 0.1 M H2SO4 (sulfuric acid.) Add 22 mL conc. sulfuric acid to 4 L deionized water. 4. Bromothymol blue and phenolphthalein pH indicator solution: 1 mg/ mL each in 4:1 isopropanol: deionized water. 5. Sodium hydroxide*, 10 M: Dissolve 40 grams sodium hydroxide in enough deionized water to make 100 mL. Do not store in glassstoppered bottle. 6. Hydrochloric acid, 0.3 M, in methanol: Dilute 2.5 mL conc. hydrochloric acid in about 80 mL methanol; dilute to 100 mL with methanol. 7. Crystal violet indicator: 2-3 mg/mL in isopropanol. 8. Reconstitution solvent: 10% acetone in toluene with 4 μg/mL of 4,4’-dibromooctafluorobiphenyl (optional).*

EQUIPMENT: NOTE: See APPENDIX B for special instructions on equipment. 1. Wipe, (7.6 cm × 7.6 cm) 12-ply or equivalent. 2. Sample storage and shipping container: 50-mL polypropylene centrifuge tubes with PTFE-lined caps. 3. Extraction tubes and vials: a. Glass test tubes, 25-mL (20 × 120 mm), with PTFE-lined caps; b. Glass test tubes, 14-mL (16 × 100 mm), with PTFE- lined caps, (ASTM Specification E982, or equivalent, suitable for repeated autoclaving); c. Amber GC autosampler vials (2-mL) and caps. 4. Gas chromatograph/mass spectrometer detector, with column and integrator, see p. 9106-1. 5. Liquid Transfer: a. Syringes: 10-, 25-, 100-, and 500-μL sizes. b. Mechanical pipette with disposable tips or repeating dispenser: 0.5-, 2.5-, and 10-mL sizes. c. Repeating dispenser: 1- to 5-mL. d. Three repeating dispensers: 10 to 20-mL each. 6. Volumetric flasks: 10-, 100-, and 250-mL. 7. Forceps. 8. Gloves, latex or nitrile. Avoid vinyl gloves (see 9106-3, Sampling step 1, NOTE 2). 9. Scoop for solid reagents. 10. Empty drying columns: 1 cm i.d. × 12-15 cm length polypropylene plastic columns having a fritted polyethylene disc or equivalent (e.g. 10-mL pipette tip with small wad of silanized glass wool packed into the tip). 11. Nitrogen blow-down apparatus with water bath capable of maintaining 35 ºC. 12. Vortex mixer. 13. Rotating mixer capable of 10-30 rpm. 14. Aspirator flask: 1-L, with aspirator tubing and a 12.5 cm long 16 gauge needle. 15. Centrifuge: capable of up to 4000 x g and of holding 25-mL glass test tubes. 16. Oven capable of 70 to 90 ºC ± 2 ºC. 17. Test tube racks, heat resistant to 90 ºC. 18. Pasteur pipettes. 19. pH paper. 20. Template: 10 cm × 10 cm or 31.7 cm × 31.7 cm opening, made of relatively rigid disposable cardstock or sheet of PTFE. 21. Ice or other cold media for shipping.

  • See SPECIAL PRECAUTIONS

Method rev. 1.1.1

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 3 of 31

SPECIAL PRECAUTIONS: The solvents are flammable and have adverse health effects. Phenethylamines target the nervous system at very low concentrations and are easily absorbed through the skin. Avoid breathing vapors. Avoid skin contact. Work should be performed in a hood with adequate ventilation. Analysts must wear proper eye and hand protection (e.g. latex gloves) to prevent absorption of even small amounts through the skin. Dissolving sodium hydroxide and concentrated hydrochloric or sulfuric acid in water is highly exothermic. Goggles must be worn. The derivatization reagents react violently with water. Caution must also be exercised in the collecting, handling, and analysis of samples. Clandestine drug labs may produce unknown and seriously toxic by-products. For example, in the manufacture of designer drugs (e.g., MPPP, a homolog of Alphaprodine), at least one very neurotoxic by-product, 1-methyl-4-phenyl-1,2,5,6-tetrahydropyridine (MPTP), has been identified that specifically and irreversibly causes Parkinson’s disease [5, 6]. SAMPLING: See APPENDIX C for special instructions on sampling. 1. Using a new pair of gloves, remove a gauze wipe from its protective package. Moisten the wipe with approximately 3 to 4 mL of methanol (or isopropanol). NOTE 1: Apply no more solvent than that needed to moisten approximately the central 80% of the area of the gauze wipe. Excess solvent may cause sample loss due to dripping from the wipe. NOTE 2: Do not use vinyl gloves due to the potential for leaching of phthalate plasticizers and contamination of the samples. 2. Place the template over the area to be sampled (may tape in place along outside edge of template). Wipe the surface to be sampled with firm pressure, using vertical S-strokes. Fold the exposed side of the pad in and wipe the area with horizontal S-strokes. Fold the pad once more and wipe the area again with vertical S strokes. 3. Fold the pad, exposed side in, and place in shipping container and seal with cap. NOTE: Keep samples refrigerated (<6 ºC). While methamphetamine and several related amines are stable on the recommended wipe media for at least 7 days at room temperature, refrigeration is recommended as soon as possible (see Table 5). 4. Either clean the template before use for the next sample or use a new disposable template. 5. Label each sample clearly with a unique sample identifier. 6. Prepare a minimum of two field blanks with one field blank for every ten samples. NOTE: In addition, include at least 3 media blanks for the analytical laboratory to use for their purposes. The wipes used for the media blanks should be from the same lot as the field samples. SAMPLE PREPARATION: See APPENDIX D for special instructions on sample preparation. 7. Desorption from media: a. Remove cap from shipping container. NOTE: Sample wipe should fit loosely in the container. If not, transfer sample to a larger container. b. Spike 60 μL of internal standard spiking solution onto each wipe sample. c. Add 30 mL desorption solution (0.1 M sulfuric acid). NOTE: If the samples were transferred to a larger container, rinse the original shipping container with the desorption solution, shake, and decant the rinsate into the larger container. d. Cap securely and mix contents by inverting the tubes end over end on a rotary mixer or equivalent at 10-30 rpm for at least one hour. NIOSH Manual of Analytical Methods (NMAM), Fifth Edition

Method rev. 1.1.1 METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 4 of 31

e. Check the pH which should be about ≤ 4. Adjust the pH with diluted (2.5 to 3 M) sulfuric acid drop-wise, mixing the contents by shaking or inversion a few times after each addition of acid before checking the pH. f. After mixing, transfer 10 mL of supernatant to a 25-mL glass centrifuge tube. NOTE: If cleanup is to be performed on a subsequent day, store samples in a refrigerator. Analytes are stable in the desorption solution for at least one week refrigerated. 8. Cleanup: Potential contamination from oils, triglycerides, plasticizers and other hydrocarbons are reduced through a hexane back-extraction step. a. To each 10-mL aliquot of acid desorbate, add 10 mL of hexane, cap and mix on a rotary mixer for one hour. Allow to stand for 15-30 minutes for the phases to separate. If an emulsion forms, centrifuge the tubes at 1500-2000 rpm for a few minutes. If the emulsion persists, add about 0.5 mL of acetonitrile to the surface of the emulsion and gently mix the layers at the interface of the emulsion. Centrifuge again if necessary. b. Aspirate the (upper) organic layer to waste. Exercise care to not remove any of the aqueous layer. 9. Extraction of analytes into methylene chloride: a. Add 1-2 drops (20-50 μL) of the mixed pH indicator (phenolphthalein + bromothymol blue) solution to each sample. The color of the samples should be yellow, which indicates that the samples are sufficiently acidic for desorption of the analytes from the wipe samples. b. Add 0.5 mL of 10 M sodium hydroxide to each sample. The color of the samples should turn brilliant purple or magenta, confirming that the pH is greater than 9-9.5 (necessary for the extraction of the amines into methylene chloride). If the color remains yellow, or only turns green or light blue, check the pH with pH paper to confirm that it is greater than 9.5. If it is not, add another 0.5 mL of 10 M sodium hydroxide, mix, and check the pH again. NOTE: The color of the solution will gradually fade from purple to deep blue within about 20-30 minutes. This is due to the known tendency of phenolphthalein to fade at high pH. c. Add 10 mL of methylene chloride to each sample. Cap and mix on a rotary mixer for one hour. Allow to stand for 15 to 30 minutes. If an emulsion forms, centrifuge as described above (step 8a). d. Aspirate the aqueous (upper) layer to waste as described above, being careful to not remove any of the lower methylene chloride layer. 10. Removal of water from the methylene chloride extract: a. Prepare potassium carbonate-sodium sulfate drying columns. Note: See APPENDIX E for preparation of drying columns. b. Rinse the packed columns with about 6 mL of methylene chloride. Dry columns afterwards by forcing dry nitrogen or clean air through the top for 10-20 seconds. c. Arrange 14-mL collection tubes (16 × 100 mm test tube) in test tube racks. Add 6 μL of crystal violet solution and 100 μL of 0.3 M hydrochloric acid in methanol to each collection tube. NOTE: Crystal violet is not critical but helps later on as a visualizing aid for monitoring the progress of drying. Hydrochloric acid is critical to prevent loss of the amphetamines during evaporative concentration. d. Position the drying columns over the collection tubes. e. Transfer (decant) the methylene chloride layer into the drying column reservoir. After the last of the sample passes into the bed of the column, rinse the drying column twice with 1 mL of methylene chloride each time and combine with sample eluate. 11. Derivatization: See APPENDIX F for special instructions on derivitization. a. Evaporate the methylene chloride eluates in a nitrogen blow-down apparatus with the water bath set at 35 °C. Rinse the tips of the evaporation needles thoroughly with methanol or acetone between samples to prevent cross-contamination. When the samples are dry, remove and cap the tubes immediately.

Method rev. 1.1.1

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 5 of 31

NOTE: The dark color of the crystal violet helps make the residue more visible when it is dried. If at least 0.1 mL of isopropanol was present in the eluates, the crystal violet will also go through a series of color changes that helps in monitoring the drying process. b. To each dried sample, add 100 μL of chlorodifluoroacetic anhydride and recap tubes. Mix the contents by vortexing briefly. NOTE 1: It is recommended that the tubes be kept capped and to only uncap about 5 at a time for the addition of the derivatizing reagent. Do not leave the acid anhydride bottle open between taking aliquots since the reagent is moisture sensitive. NOTE 2: If incomplete derivatization is routinely experienced, increase volume of reagent to 150 or 200 μL. The color of the crystal violet will turn yellow or yellow-green with the addition of chlorodifluoroacetic anhydride. c. Heat in an oven at 70-75 °C for 20-30 minutes. d. After heating, allow the tubes to cool to room temperature. Remove caps and evaporate the contents to dryness under a stream of nitrogen at room temperature. As the solution concentrates it turns from a yellow or yellowish-green solution to a bluish-green just before going to dryness. At the point of dryness the color of the residue normally turns rapidly to blue or violet, depending upon the amount of coextractants (the more co-extractants, the more blue the color and the less likely a violet color will develop). Remove the tubes just as soon as the blue or violet color becomes apparent. Losses have been experienced if blowing is continued for more than 2 minutes beyond the blue or violet color stage. NOTE: If an oil-like residue or film persists, then the sample may have too many contaminants that were not removed at the cleanup step or were introduced subsequent to cleanup. In such a case, return to step 7f and perform the clean-up (step 8) on another 10-mL aliquot of the sample desorbate using methylene chloride as the cleanup solvent instead of hexane. Discard the (lower) organic layer to waste before proceeding to steps 9 through 11. e. Reconstitute the dried residue with 1 mL of the reconstitution solvent. The reconstituted solution normally will become deep blue in color. Mix by vortexing briefly a couple of times. Transfer the solutions to 2-mL amber-colored GC vials containing 200 to 250 mg anhydrous sodium sulfate. Cap vials, label, and analyze by GC/MS (See MEASUREMENT, steps 15-17). NOTE: Derivatives of phenylpropanolamine (norephedrine) break down significantly over several days at room temperature. GC vials containing derivatives should be kept refrigerated until analysis. CALIBRATION AND QUALITY CONTROL: 12. Determine retention times for the derivatives of the analytes of interest using the column and chromatographic conditions specified on page 9106-1. Table 10 gives typical retention times for various drugs, precursors, and adulterants. 13. Calibrate daily with at least six calibration standards plus a blank (CS0) selected from Table 7 to cover the analytical range. a. Prepare the analyte spiking solution as follows: Add known amounts of individual drug stock solutions to a volumetric flask and dilute to volume with methanol. A recommended final concentration for this solution is approximately 200 μg each per mL. b. Prepare calibration standards and media blanks in clean shipping containers (e.g. 50-mL polypropylene centrifuge tubes or equivalent). NOTE: Liquid standards (standards without added blank wipe media) may be prepared in lieu of media standards if cotton gauze was used for the samples. c. Add 3 mL methanol (or isopropanol if isopropanol was used with the samples in the field) to each calibration standard and media blank.

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition

Method rev. 1.1.1 METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 6 of 31

d. Spike a known volume of analyte spiking solution into each calibration standard by spiking directly onto the media or into solution. Use the spiking volumes suggested in Table 7 to cover the desired range. e. Process each of these through steps 7 through 11 (same as the field samples.) f. Analyze these along with the field samples. (See MEASUREMENT, steps 15-17.) 14. Prepare matrix-spiked (QC) and matrix-spiked duplicate (QD) quality control samples [7]. a. Cotton gauze from the same lot used for taking samples in the field should be provided to the analytical laboratory to prepare these matrix-spiked quality control samples. b. The quality control samples (QC and QD) must be prepared independently at concentrations within the analytical range. (See Table 7 for applicable concentration ranges.) c. One quality control media blank (QB) must be included with each QC and QD pair. i.  Transfer clean gauze wipes to new shipping containers. ii.  Add 3 mL of isopropanol (or methanol if methanol was used in wiping) to each gauze wipe. iii. Spike QC and QD with a known amount of analyte as suggested in Table 7. d. Process each of these through steps 7 through 11 (same as the field samples). e. Analyze these along with the field samples. (See MEASUREMENT, steps 15-17.) MEASUREMENT: See APPENDIX G for special instructions on measurement. 15. Analyze the calibration standards, quality control samples, blanks, and samples by GC/MS. a. Set gas chromatograph according to manufacturer’s recommendations and to conditions listed on page 9106-1. b. Set mass spectrometer conditions to manufacturer’s specifications and those given on page 9106-1 for the scan mode or those in Table 6 for the SIM mode. c. Inject sample aliquot with autosampler or manually. NOTE: After the derivatives are prepared and just before analyzing any samples or standards, inject the highest concentrated standard several times in order to prime or deactivate the GC column and injection port. This will help minimize any drift in the instrument’s response to target analytes relative to their internal standards. d. After analysis, the vials should be recapped promptly and refrigerated if further analysis is anticipated. 16. Using extracted ion current profiles for the primary (quantification) ions specific to each analyte, measure GC peak areas of analyte(s) and internal standard(s) and compute relative peak areas by dividing the peak area of the analyte by the area of the appropriate internal standard. Recommended primary (quantification) ions and internal standards are given in Tables 6, 8, and 9. Prepare calibration graph (relative peak area vs. μg analyte per sample). 17. Samples from initial investigations of clandestine laboratories are likely to include highly contaminated samples. If sample results exceed the upper range of the calibration curve, the sample in the GC vial may be diluted and reanalyzed or a smaller aliquot of the initial acid desorbate diluted, re-extracted, derivatized, and analyzed. Refer to APPENDIX H for instructions and limitations on making dilutions. CALCULATIONS: 18. Determine the mass (in μg/sample) of respective analyte found in the wipe samples, and in the media blank from the calibration graph. 19. Calculate final concentration, C, of analyte in μg/sample:

= C c

V V1 V3 −b 5 V2 V4 V2

Where: c = concentration in sample (in μg/sample determined from the calibration curve). Method rev. 1.1.1

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 7 of 31

V1 = volume correction factor (needed only when the volume of internal standard V2 spiking solution used for spiking the samples – such as for composite samples requiring larger desorption solution volumes – is different from that used for spiking the calibration standards). (See Table 7, footnote 4.)

V1 =  volume in μL of internal standard spiking solution used to spike samples. V2 =  volume in μL of internal standard spiking solution used to spike the standards.

V3 = dilution factor, if applicable V4 V3 = 10 mL (volume of desorbate taken for cleanup in step 8). V4 = volume in mL of desorbate actually taken for cleanup and diluted to 10 mL with blank desorbing solution containing internal standard. b = concentration in media blank (in μg/sample determined from the calibration curve). V5 = volume correction factor for the media blank (needed only if the volume of V2 internal standard spiking solution used for spiking the media blank is different from that used for spiking the calibration standards.) V5 =   volume in μL of internal standard spiking solution used to spike media blank. 20. Report concentration, C’, in μg per total area wiped (in cm2) as follows:

C’=

C A

Where: C = μg/sample (step 19). A = Total area wiped in cm2 per sample. NOTE: For example, if the sample was a composite sample and the area was 400 cm2, report results as μg/400 cm2 and not averaged to μg/100 cm2. In general, if the area wiped was greater than or less than 100 cm2, do not convert value to μg/100 cm2. To avoid confusion, report separately both μg/sample (C) and the total area wiped in cm2 per sample (A) for both discrete and composite samples. EVALUATION OF METHOD: This method was evaluated for those analytes listed in Tables 8a and 8b over a range of approximately 0.1 μg/sample to 30 μg/sample. These concentration levels represent approximately the 1 through 300 times the limit of quantitation (LOQ) level for most of the analytes [8]. Results are reported in the Backup Data Report for NIOSH 9106 [1]. The limits of detection (LOD and LOQ) were determined by preparing a series of liquid standards in desorption solution, processing them through the liquid-liquid extraction procedure of NIOSH 9106, and analyzing in both the scan and SIM modes. The LODs were estimated using the procedure of Burkart [8]. An LOD of 0.05 μg/sample for methamphetamine on wipes was achieved in either scan or SIM mode. The LOD was set at 0.05 μg/sample because that was the level of the lowest calibration standard for the LOD study. Lower LODs (e.g. 0.02 μg/sample) have been achieved in practice by including calibration standards at lower concentration levels. The cleanliness and performance of the mass spectrometer must be maintained such that at 0.1 μg/sample, a signal of at least 5 to 10 times the baseline noise is achievable. This is more easily accomplished in the SIM mode with the mass spectrometer. Six different wipe media were evaluated. These were 3”×3” 12-ply cotton gauze, 4”×4” AlphaWipes® (TX® 1004), 4”×4” 4-ply NU GAUZE®, 4”×4” 4-ply MIRASORB®, 4”×4” 6-ply SOF-WICK®, and 4”×4” 4-ply TOPPER® sponges. Results are given in the Backup Data Report [1]. No synthetic media performed NIOSH Manual of Analytical Methods (NMAM), Fifth Edition

Method rev. 1.1.1 METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 8 of 31

better than cotton gauze. Some media (TOPPER® and SOF-WICK®) did not perform well, possibly due to co-extracted nonionic (polyethoxyethylene type) surfactants that are not removed using hexane and incompletely removed using methylene chloride in the cleanup step. Precision and accuracy were determined by analyzing 6 replicates at each of 6 concentration levels (nominally 0.1, 0.3, 1, 3, 10, and 30 μg/sample). Results are presented in Tables 8a and 8b for cotton gauze. The best precision and accuracies were dependent upon the use of carefully chosen internal standards, especially where there is steric hindrance of the amine (e.g. having N-ethyl and N-propyl groups). Long term sample storage stability was determined for periods up to 30 days under refrigeration (4 °C ± 2 °C) and for up to 7 days at room temperature (22-24 °C). Results are given in Table 5. The precision and accuracy and long term storage stability evaluations were conducted using isopropanol as the wetting solvent. A second precision and accuracy study using methanol confirmed that methanol is an acceptable substitute for isopropanol. Recovery of amphetamines from six different types of surfaces using cotton gauze was evaluated (see Tables 11a and 11b). The practice of serial wiping (wiping the same surface area a second time with a second gauze wipe and combining both wipes as a single sample) was evaluated. Four solvents for wetting the gauze were tested (distilled water, 5% distilled white vinegar, isopropanol, and methanol). Six replicate samples were made on a latex painted wall. Recoveries and precisions are given in Table 11a. The recoveries with 5% distilled white vinegar were better than for distilled water, but not as good as for isopropanol. Methanol is superior to isopropanol. Recoveries with isopropanol are greatly improved with a repeat (serial) wipe (11% improvement compared to only about 6% improvement with methanol). The study and results are reported in the Backup Data Report for NIOSH 9109 [9]. Additional research on surface sample recovery and solvent effectiveness has been reported by Martyny [11, 12]. REFERENCES: [1] Reynolds JM, Siso MC, Perkins JB [2004]. “Backup data report for NIOSH 9106, Methamphetamine and illicit drugs, precursors, and adulterants on wipes by liquid-liquid extraction, Abridged,” prepared under NIOSH Contract 200-2001-0800, (Unpublished, 2004). [2] Martyny JW, Arbuckle SL, McCammon CS Jr., Esswein EJ, Erb N [2003]. Chemical exposures associated with clandestine methamphetamine laboratories. http://www.njc.org/pdf/chemical_ exposures.pdf [2005]. [3] NIOSH [2011]. Methamphetamine and illicit drugs, precursors, and adulterants on wipes by solid phase extraction: Method 9109. In: Ashley KA, O’Connor PF, eds. NIOSH manual of analytical methods. 5th ed. Cincinnati, OH: U.S. Department of Health and Human Services, Centers for Disease Control and Prevention, National Institute for Occupational Safety and Health, [www.cdc. gov/niosh/nmam/]. [4] NIOSH [2011]. Methamphetamine on Wipes by Liquid Chromatography/Mass Spectrometry: Method 9111. In: Ashley KA, O’Connor PF, eds. NIOSH manual of analytical methods. 5th ed. Cincinnati, OH: U.S. Department of Health and Human Services, Centers for Disease Control and Prevention, National Institute for Occupational Safety and Health, [www.cdc.gov/niosh/nmam/]. [5] Baum RM [1985]. New variety of street drugs poses growing problem. Chemical and Engineering News, 63(63):7-16. [6] Buchanan JF, Brown CR [1988]. Designer drugs, a problem in clinical toxicology. Medical Toxicology 3:1-17. [7] NIOSH [1994]. Chapter C: Quality Assurance. In: Eller PM, Cassinelli ME, eds. NIOSH Manual of Analytical Methods, 4th ed. Cincinnati, OH: Department of Health and Human Services, Centers for Disease Control and Prevention, National Institute for Occupational Safety and Health, DHHS (NIOSH) Publication No. 94-113. [www.cdc.gov/niosh/nmam/]. [8] Burkart JA [1986]. General procedures for limit of detection calculations in the industrial hygiene chemistry laboratory. Applied Industrial Hygiene 1(3):153-155. Method rev. 1.1.1

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 9 of 31

[9] Reynolds JM, Siso MC, Perkins JB [2004]. Backup data report for NIOSH 9109, Methamphetamine and illicit drugs, precursors, and adulterants on wipes by solid phase extraction. Unpublished. Prepared under NIOSH Contract 200-2001-0800. [10] Merck Index, 11th ed., S. Budavari, Ed., Merck and Co., Rahway, NJ (1989). [11] Martyny JW. Decontamination of building materials contaminated with methamphetamine [http://health.utah.gov/meth/html/Decontamination/ DecontaminationofBuildingMaterialsContaminatedwithMeth.pdf ]. Date accessed: May, 2007. [12] Martyny JW [2008]. Methamphetamine sampling variability on different surfaces using different solvents [http://health.utah.gov/meth/html/Decontamination/ MethSamplingVariabilityonDifferentSurfaces.pdf ] Date accessed: May, 2011. [13] Sweet DV, ed. [1997]. Registry of toxic effects of chemical substances. DHHS (NIOSH) Publ. No. 97-119. http://www.cdc.gov/niosh/pdf/97-119.pdf. [14] Sigma-Aldrich MSDS sheets, http://www.sigma-aldrich.com [May 10, 2004]. [15] Cerilliant Analytical Reference Materials. 811 Paloma Drive, Suite A, Round Rock, Texas 78664 (www. cerilliant.com). [16] NAMSLD 2007. State Controlled Substance(s) Environmental Issues Bill – Status Update, ( http:// www.natlalliance.org/ ) The National Alliance for Model State Drug Laws, Alexandra, Va. Access via the web on March 11, 2008. [17] Howard PH, Meylan WM [1997]. Handbook of physical properties of organic chemicals. Boca Raton, Florida: CRC/Lewis Publishers. See also Syracuse Research Corporation’s website http://www.syrres. com/esc/physdemo.htm (May 10, 2004) for demos and updates. [18] Zhang JS, Tian Z, Lou ZC [1989]. Quality evaluation of twelve species of chinese ephedra (Ma Huang). Acta Pharmaceutic Sinica 24(11):865-871. [19] Cui JF, Zhou TH, Zhang JS, and Lou ZC. [1991]. Analysis of alkaloids in chinese ephedra species by gas chromatographic methods. Phytochemical Analysis 2:116-119. [20] Andrews KM [1995]. Ephedra’s role as a precursor in the clandestine manufacture of methamphetamine. Journal of Forensic Sciences 40(4):551-560. [21] OSHA [2001]. Evaluation guidelines for surface sampling methods. Industrial Hygiene Chemistry Division, OSHA Salt Lake Technical Center, Salt Lake City, Utah. [http://www.osha.gov/dts/sltc/ methods/surfacesampling/surfacesampling.pdf ] Date accessed: May 2011. METHOD DEVELOPMENT BY: John M. Reynolds, Maria Carolina Siso, and James B. Perkins, DataChem Laboratories, Inc., Salt Lake City, Utah under NIOSH Contract CDC 200-2001-0800.

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition

Method rev. 1.1.1 METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 10 of 31

Table 1. Formula and registry numbers of analytes. MW(1) (Daltons) Compound (alphabetically) (DL)-Amphetamine

Free base 135.21

Hemisulfate Structural Formula HCl salt salt As free base 171.67 184.25 C6H5·CH2·CH(CH3)·NH2

(D)-Amphetamine(7)

135.21

171.67

184.25

C6H5·CH2·CH(CH3)·NH2

(L)-Amphetamine

135.21

171.67

184.25

Caffeine

194.19

(DL)-Ephedrine

165.24

201.70

(L)-Ephedrine(8)

165.24

(D)-Ephedrine

RTECS(6) SH9450000 SI1750000 SI1400000

C6H5·CH2·CH(CH3)·NH2

CAS #(2) 300-62-9(3) 60-13-9(5) 51-64-9(3) 51-63-8(5) 156-34-3(3)

(CH3)3·[C5HN4O2]

58-08-2

EV6475000

214.28

C6H5·CH(OH)·CH(CH3)·NH·CH3

201.70

214.28

C6H5·CH(OH)·CH(CH3)NH·CH3

90-81-3(3) 134-71-4(4) 299-42-3(3) 50-98-6(4) 134-72-5(5)

165.24

201.70

214.28

C6H5·CH(OH)·CH(CH3)NH·CH3

(±)-MDEA

207.27

243.73

CH2O2C6H3·CH2·CH(CH3)NH·C2H5

(±)-MDMA

193.24

229.71

CH2O2C6H3·CH2·CH(CH3)·NH·CH3

(+)-MDMA(7) 193.24 (DL)-Methamphetamine 149.24

229.71 185.70

198.28

CH2O2C6H3·CH2·CH(CH3)·NH·CH3 C6H5·CH2·CH(CH3)·NH·CH3

(D)-Methamphetamine(7) 149.24

185.70

198.28

C6H5·CH2·CH(CH3)·NH·CH3

(L)-Methamphetamine

149.24

185.70

198.28

C6H5·CH2·CH(CH3)·NH·CH3

Phencyclidine

243.39

279.85

C6H5·C[C5H10]·N[C5H10]

Phentermine

149.24

185.70

C6H5·CH2·C(CH3)2·NH2

(DL)-Norephedrine

151.21

187.67

200.25

C6H5·CH(OH)·CH(CH3)·NH2

1R,2S (-)-Norephedrine

151.21

187.67

200.25

1S,2R (+)-Norephedrine 151.21

187.67

1S,2S (+)-Norephedrine 151.21

(3)

321-98-2(3) 24221-86-1(4) 82801-81-8(3) 116261-63-2(4) 42542-10-9(3) 92279-84-0(4) 64057-70-1(4) 4846-07-5(3)

SH9050000

KB0700000 KB1750000 KB2625000 KB0600000 KB1925000

SH5700000 SH5700000

537-46-2(3) 51-57-0(4) 33817-09-3(3)

SH4910000 SH5455000 SH4905000 TN2272600 TN2272600 SH4950000

C6H5·CH(OH)·CH(CH3)·NH2

77-10-1(3) 956-90-1(4) 122-09-8(3) 1197-21-3(4) 14838-15-4(3) 154-41-6(4) 492-41-1(3)

200.25

C6H5·CH(OH)·CH(CH3)·NH2

37577-28-9(3)

187.67

200.25

C6H5·CH(OH)·CH(CH3)·NH2

(D)-Pseudoephedrine(8,9) 165.24

201.70

214.28

C6H5·CH(OH)·CH(CH3)·NHCH3

(L)-Pseudoephedrine(10)

201.70

214.28

C6H5·CH(OH)·CH(CH3)·NH·CH3

36393-56-3 2153-98-2(4) 492-39-7(4) 90-82-4(3) 345-78-8(4) 321-97-1(3)

165.24

RC2625000 DN4200000 RC2275000

RC9275000

UL5800000 UL5950000

(1) Molecular weights are calculated from the empirical formula using the 1987 IUPAC Atomic Weights of the Elements, Merck Index [9]. The molecular weight of the hemisulfate is ½ the weight of the 2:1 sulfate salt (2 moles amine + 1 mole H2SO4). (2) CAS from various sources: Merck Index [10], NIOSH RTECS [13], MSDS sheets from Sigma/Aldrich [14], Cerilliant [15], and other sources [18]. (3) Free base form. (4) Hydrochloride salt. (5) 2:1 Sulfate salt (2 moles amine + 1 mole H2SO4. (6) RTECS = NIOSH Registry of Toxic Effects of Chemical Substances [13]. (7) More active isomer. (8) Naturally occurring isomer. (9) The D form of pseudoephedrine is a decongestant. (10) The L form of pseudoephedrine is a bronchodilator. Dehydroxylation forms the less active L-methamphetamine. Method rev. 1.1.1

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 11 of 31

Table 2. Methamphetamine Regulations by State (Jan 2008)* State

Standard

Alaska**

0.1 μg/100 cm

Arizona

State

Standard

Minnesota

0.1 μg/100 cm2 (meth labs), < 1.5 μg/100 cm2 (meth use)

0.1 μg/100 cm2

Montana

0.5 μg/ft2

Arkansas

0.1 μg/100 cm2

New Mexico

1.0 μg/ft2

California***

< 1.5 μg/100 cm2

North Carolina

Colorado

0.5 μg/100 cm2

Oregon

Connecticut

0.1 μg/100 cm2

South Dakota

0.1 μg/100 cm2

Hawaii

0.1 μg/100 cm2

Tennessee

0.1 μg/100 cm2

Idaho

0.1 μg/100 cm2

Utah

0.1 μg/100 cm2

Kentucky

0.1 μg/100 cm

Washington

<0.1 μg/100 cm2

2

2

0.1 μg/100 cm2 0.5 μg/ft2

The following states have no standard: Alabama, Delaware, D.C., Florida, Georgia, Illinois, Indiana, Iowa, Kansas, Louisiana, Maine, Maryland, Massachusetts, Michigan, Mississippi, Missouri, Nebraska, Nevada, New Hampshire, New Jersey, New York, North Dakota, Ohio, Oklahoma, Pennsylvania, Rhode Island, South Carolina, Texas, Vermont, Virginia, West Virginia, Wisconsin, Wyoming. NIOSH has not established health-based or feasibility-based airborne Recommended Exposure Limits (RELs) or surface contamination guidelines for clandestine drug laboratories. State surface contamination limits are provided as an aid to those seeking additional information and does not constitute endorsement by NIOSH. The National Alliance for Model State Drug Laws (NAMSDL) website: http://www.namsdl.org/home.htm periodically summarizes state feasibility-based decontamination limits and proposed state legislative requirements and guidelines [16]. However, state information is subject to change, and specific state’s surface contamination limits, and other state decontamination requirements and guidelines should be obtained directly from each state.

    • Guidance and Standards for Cleanup of Illegal Drug-Manufacturing Sites Revision 1 April 19, 2007 Alaska Department of Environmental Conservation, Spill

Prevention and Response Division, Prevention and Emergency Response Program. http://www.dec.alaska.gov/spar/perp/methlab/druglab_guidance.pdf

      • In Oct 2009 House Bill 1489 was passed into law to incorporate the new standard as the state limit. All other states: Data source: http://health.utah.gov/meth/

html/Resources/OtherStates/Nationalcomparison (downloaded April 2011).

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition

Method rev. 1.1.1 METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 12 of 31

Table 3. Physical properties of analytes(1) Compound (alphabetically)

CAS

m.p.(°C)

Vapor Pressure (mm Hg)

pKa(4) 10.1 @ 20 °C

Log P(5)

Solubility in Water, g/100 mL 2.8 @ 25 °C

(DL)-Amphetamine

300-62-9

(D)-Amphetamine

51-64-9

<25

9.9

1.76

(D)-Amphetamine sulfate

51-63-8

>300

6.81

(L)-Amphetamine

156-34-3

0.201 @ 25 °C

10.1 @ 20 °C

1.76

2.8 @ 25 °C

Caffeine

58-08-2

238

15 @ 89 °C

10.4 @ 40 °C

-0.07

2.16 @ 25 °C

(DL)-Ephedrine

90-81-3

76.5

0.68

(L)-Ephedrine

299-42-3

34

0.00083 @ 25 °C

10.3 @ 0 °C

1.13

6.36 @ 30 °C

(L)-Ephedrine HCl

50-98-6

218

2.04E-10 @ 25 °C

pH 5.9 @ 1/200 dil.(3)

-2.45

25(6)

MDEA

82801-81-8

MDMA HCl

42542-10-9

148-149(2)

(D)-Methamphetamine

537-46-2

0.163 @ 25 °C

9.87 @ 25 °C

2.07

1.33 @ 25 °C

(D)-Methamphetamine HCl

51-57-0

170-175(2)

Phencyclidine

77-10-1

46.5

8.29

4.69

Phencyclidine HCl

956-90-1

233-235(2)

Phentermine

122-09-8

0.0961 @ 25 °C

1.90

1.86 @ 25 °C

Phentermine HCl

1197-21-3

198(2)

(±) Phenylpropanolamine

14838-15-4

0.000867 @ 25 °C

9.44 @ 20 °C

0.67

14.9 @ 25 °C

(±) Phenylpropanolamine HCl

154-41-6

194

-2.75

(L)-Norephedrine

492-41-1

51-53

1S,2S (+)-Norephedrine

36393-56-3

77.5-78

0.000867 @ 25 °C

0.83

14.9 @ 25

1S,2S (+)-Norephedrine HCl

492-39-7

0.22

2 @ 25

(D)-Pseudoephedrine

90-82-4

119

0.00083 @ 25 °C

9.44 @ 20 pH 5.9-6.1 in aq. soln. (3) 10.3 @ 0 °C

0.89

10.6 @ 25 °C

(D)-Pseudoephedrine HCl

345-78-8

181-182(2)

pH 5.9 @ 1/200 dil.(3)

(1) (2) (3) (4) (5) (6)

(3)

1.76

Handbook of Physical Properties of Organic Chemicals unless otherwise noted [17]. Merck Index [10]. Sigma-Aldrich MSDS [14]. Negative log of the acid dissociation constant for the amine in aqueous solution. Log P = octanol-water partition coefficient. Temperature not given in source.

Method rev. 1.1.1

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 13 of 31

Table 4. Synonyms of analytes Generic names(1) (DL)-Amphetamine; (±)-Amphetamine (D)-Amphetamine; (+)-Amphetamine (L)-Amphetamine; (-)-Amphetamine Caffeine (DL)-Ephedrine; (±)-Ephedrine (L)-Ephedrine; (-)-Ephedrine; (1R,2S)-(-)-Ephedrine; l-Ephedrine (D)-Ephedrine MDEA MDMA (DL)-Methamphetamine; (±)-Methamphetamine (D)-Methamphetamine; (+)-Methamphetamine; d-Methamphetamine (L)-Methamphetamine; (-)-Methamphetamine Phencyclidine Phentermine (DL)-Norephedrine; (±)-Norephedrine (L)-Norephedrine; (-)-Norephedrine (D)-Norephedrine; (+)-Norephedrine (+)-Norpseudoephedrine; Cathine L-(+)-Pseudoephedrine; (+)-Pseudoephedrine; d-Pseudoephedrine D-(-)-Pseudoephedrine; (-)-Pseudoephedrine

Trade and street names(2) Benzedrine; Phenedrine; Bennies

Additional names(3) (±)-α-Methylbenzeneethanamine(4); dl-α-Methylphenethylamine(4); dl-1-Phenyl-2aminopropane; (±)-Desoxynorephedrine Dextroamphetamine; Dexedrine; (S)-α-Methylbenzeneethanamine(4); d-α-Methylphenethylamine(4); dexies d-1-phenyl-2-aminopropane; d-β-Phenylisopropylamine Levoamphetamine; component of (R)-α-Methylbenzeneethanamine(4); l-α-Methylphenethylamine(4); Adderall l-1-phenyl-2-aminopropane; (-)-1-phenyl-2-aminopropane Component (with ephedrine) of 3,7-Dihydro-1,3,7-trimethyl-1H-purine-2,6-dione(4); cloud 9 and herbal XTC 1,3,7-Trimethylxanthine Ephedral; Racephedrine; Sanedrine (R*,S*)-(±)-alpha-[2-(Methylamino)ethyl]benzenemethanol; DL-alpha-[1-(Methylamino)ethyl]benzyl alcohol; dl-Ephedrine Primatene; Xenadrine; Ma Huang (R-(R*,S*))-α-(1-Methylaminoethyl)benzenemethanol; L-erythro-2(Ephedra sinica and other species(5)); (Methylamino)-1-phenylpropan-1-ol; (1R,2S)-(-)-2-Methylamino-1(with caffeine) cloud 9 and herbal phenyl-1-propanol; (-)-alpha-(1-Methylamino-ethyl)-benzyl alcohol; ecstasy (-)-1-hydroxy-2-methylamino-1-phenylpropane; L-(-)-Ephedrine (1S,2R)-(+)-2-Methylamino-1-phenyl-1-propanol; (+)-Ephedrine MDE; Eve (±)-3,4-Methylenedioxy-N-ethylamphetamine; N-ethyl-alpha-methyl-1,3-benzodioxole-5-ethanamine Adam, ecstasy, X, XTC N,α-Dimethyl-3,4-1,3-benzodioxole-5-ethanamine; 3,4-Methylenedioxymethamphetamine N,α-Dimethylbenzeneethanamine(4); N,α-Dimethylphenethylamine; dl-Desoxyephedrine; N-methyl-β-phenylisopropylamine Methedrine; Desoxyn; chalk; crank; (S)-N,α-Dimethylbenzeneethanamine; (S)-(+)-N,α-Dimethylcrystal; glass; ice; meth, speed; upper phenethylamine(4); d-1-Phenyl-2-methylaminopropane; d-Desoxyephedrine; d-N-methyl-β-phenyl-isopropylamine Component in decongestant vapor (R)-(-)-N,α-Dimethylphenethylamine; (-)-Deoxyephedrine; inhaler (Vick’s brand) (-)-2-(Methylamino)-1-phenylpropane Sernylan; Sernyl; angel dust; PCP; 1-(1-Phenylcyclohexyl) piperidine(4) peace pill Fastin; Normephentermine α,α-Dimethylbenzeneethanamine(4); α,α-Dimethylphenethylamine(4); 1,1-Dimethyl-2-phenylethylamine; α-Benzylisopropylamine (±)-Phenylpropanolamine; Obestat; (R*,S*)-(±)-α-(1-Aminoethyl)benzenemethanol(4); -(±)-α-(1-AminoPhenedrine; ethyl)benzyl alcohol(4); (±)-2-Amino-1-phenyl-1-propanol Natural form found in Ephedra sinica (1R,2S)- 2-Amino-1-phenyl-1-propanol; (1R,2S)-Norephedrine; and other species(5) l-erythro-2-Amino-1-phenylpropan-1-ol Metabolite of cathinone in urine of (1S,2R)- 2-Amino-1-phenyl-1-propanol; (1S,2R)-Norephedrine; Khat users. d-erythro-2-Amino-1-phenylpropan-1-ol Amorphan; Adiposettin; Reduform; (R*,R*)-α-(1-Aminoethyl)benzenemethanol(4); d-threo-α-2-Amino-1found naturally in Khat plant hydroxy-1-phenylpropane; 1S,2S-(+)-Norpseudoephedrine Afrinol; Novafed; Sinufed; Sudafed; (S-(R*,R*))- α-[1-(Methylamino)ethyl]benzenemethanol; (1S,2S)natural form found in Ephedra sinica (+)-2-Methylamino-1-phenylpropanol; d-(alpha-(1-Methylamino)and other species(5) ethyl)benzyl alcohol; (1S,2S)-(+)-Pseudoephedrine; d-threo-2Methylamino-1-phenylpropan-1-ol; (+)-ψ-Ephedrine (1R,2R)-(-)-Pseudoephedrine; (-)-ψ-Ephedrine; l-threo-2Methylamino-1-phenylpropan-1-ol; (+)-ψ-Ephedrine

(1) Common or generic names. Salts forms are not given for simplicity. (2) Trade and street names are exemplary, not exhaustive. Street names change over time and by locality. Salts and free base forms are not distinguished. (3) Other names from Merck Index [10], NIOSH Registry of Toxic Effects of Chemical Substances [13], and MSDS sheets [14] and other reference materials [15]. NOTE: For amphetamine and methamphetamine the prefixes R-, D-, d-, and (+)- , although they mean different things, are essentially synonymous for the dextrorotatory stereoisomer and S-, L-, l-, and (-) are essentially synonymous for the levorotary stereoisomer. Many other synonyms exist. (4) Uninverted CAS name as given in Merck Index [10]. (5) Extracts of Ephedra species contain various amounts of (+)-Norephedrine, (-)-N-methylephedrine, and (+)-N-methylpseudoephedrine. (+)-Norephedrine is reduced to amphetamine and N-methylephedrine and N-methylpseudoephedrine reduce to N,N-dimethylamphetamine [18, 19]. The presence of these latter two compounds in methamphetamine samples indicate that Ephedra spp. extracts may have been used in the synthesis [20].

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition

Method rev. 1.1.1 METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 14 of 31

Table 5. Limit of detection (LOD), method detection limit (MDL), and sample storage stability on cotton   gauze. (1) Estimated LOD(3)

Compound

Int. std.(2)

Estimated MDL(4)

Scan Mode SIM Mode Scan Mode (µg/sample) (µg/sample) (µg/sample)

Storage Stability(5)

SIM Mode 30 days 4 °C (µg/sample)

7 days 22 °C

(D)-Amphetamine

D11-Amp D14-Met

0.07 0.06

0.05 0.06

0.04 0.03

0.02 0.03

100.5 99.7

94.5 87.9

Caffeine

D11-Amp D14-Met

1 1

0.2 0.2

0.4 (6) 0.4 (6))

0.02 0.03

99.3 98.5

98.8 91.9

(L)-Ephedrine

D11-Amp D14-Met

0.09 0.08

0.1 0.09

0.02 0.01

0.01 0.06

95.6 94.8

97.2 90.5

MDEA

N-PAmp

0.05

0.07

0.1

0.02

98.9

102.1

MDMA

D11-Amp D14-Met

0.05 0.05

0.06 0.07

0.04 0.03

0.02 0.02

99.7 98.9

111.1 103.2

(D)-Methamphetamine

D11-Amp D14-Met

0.07 0.05

0.05 0.05

0.03 0.03

0.02 0.02

98.7 98.0

100.6 93.5

Phencyclidine

D11-Amp D14-Met

0.3 0.3

0.06 0.07

0.03 0.03

0.02 0.02

103.7 102.9

105.2 97.7

Phentermine

D11-Amp D14-Met

0.06 0.05 0.06 0.2 0.2 0.08 0.07 0.06

0.05 0.05 0.05

0.02 0.02 0.01 0.1 (9) 0.2 (10) 0.03 0.05 0.05

0.02 0.02

102.0 101.1

101.5 94.3

(8)

94.3 93.6 100.4 99.6 -

92.7 86.2 97.9 91.1 -

(±)-Norephedrine(7) Pseudoephedrine

D11-Amp D14-Met D11-Amp D14-Met NMPhen

(8)

0.07 0.09 0.09

(8)

0.02 0.03 0.02

(1) Backup Data Report [1]. (2) Internal standards: D11-Amp = Amphetamine-D11, D14-Met = Methamphetamine-D14, NMPhen = N-Methyl phenethylamine, N-PAmp = N-Propyl amphetamine. (3) LODs vary according to individual GC columns, instrument conditions and cleanliness, media interferences, and internal standards used. The lowest calibration standard for these determinations was 0.05 μg/sample. Lower LODs are achievable with lower concentration calibration standards and operation of the mass spectrometer in the SIM mode. LODs were calculated on liquid standards using the procedure of Burkart [8]. (4) MDLs are provided as an alternate expression of sensitivity. These MDLs are calculated as the standard deviation of six replicates on spiked media analyzed at the 0.1 μg/sample level (except as noted) times the Student’s t value for 6 replicates (3.365). (Normally 7 replicates are used.) (5) Cotton gauze samples were spiked at 3 μg/sample per analyte. Six samples were analyzed immediately after preparation. Six samples were stored at room temperature (about 22 °C) for 7 days and then analyzed. Eighteen samples were stored at +4 °C (±2 °C). Of the 18 samples stored at +4 °C, six each were analyzed at 7 and 21 days and three each were analyzed at 14 and 30 days. (Backup Data Report [1].) Apparent recoveries vary according to internal standard used. (6) The 0.3 μg/sample level was undetectable in the scan mode. MDLs were calculated from the 1 μg/sample level. (7) (±)-Norephedrine = (±)-phenylpropanolamine. (8) (±)-Norephedrine was not evaluated in the SIM mode due to breakdown of derivative with room temperature storage for one week. (9) MDL calculated from the 0.3 μg/sample level. (Recoveries were >120% at the 0.1 μg/sample level.) (10) MDL calculated from the 1 μg/sample level. (Recoveries were >120% at the 0.1 and 0.3 μg/sample levels.)

Method rev. 1.1.1

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 15 of 31

Table 6. Example of mass spectrometer operation parameters for selected ion monitoring mode.(1) Chlorodifluoroacetyl derivatives Scan window(2) Acquisition ions (m/z) per group(3) Acquisition Group 1 10.5 to 13.0 104 118 128 156 160 170 172 177 Acquisition Group 2 13.0 to 15.2 104 156 158 170 172 198 296 Acquisition Group 3 15.2 to 18.0 109 135 162 170 184 194 200 242 (6) (7) GC Peak Retention Time Primary Ion (m/z) Secondary ion and approximate relative abundance(8) (4) (5) No. Target Analytes and Internal Standards: (min) (Quantification Ion) (relative to the Primary Ion) Acquisition Group 1 85% 2 Amphetamine-D11 (I$)(9) 11.07 160 128 85% 3 Amphetamine 11.15 156 118 33% 5 Phentermine 11.34 170 172 95% 8 n-Methyl phenethylamine (I$)(9) 12.20 156 104 32% 9 Methamphetamine-D14 (I$)(9) 12.51 177 128 32% 10 Methamphetamine 12.61 170 118 Acquisition Group 2: 18 Phenylpropanolamine 13.27 156 246 25% 20 Dibromooctafluorobiphenyl(10) 13.63 296 456 115% 21 N-n-Propylamphetamine (I$)(9) 13.8 198 156 75% 25 Ephedrine 14.27 170 172 33% 28 Pseudoephedrine 14.74 170 172 33% Acquisition Group 3: 32 Caffeine 15.66 194 109 50% 40 Phencyclidine 16.41 200 242 35% 41 MDMA 16.48 170 162 95% 43 MDEA 16.87 184 162 75% (1) In this example, 10 analytes and 5 internal standards are grouped into 3 acquisition groups having no more than 8 primary and secondary ions per acquisition group. For 6 analytes and internal standards or less, one acquisition group may be sufficient. (2) Scan window is in minutes. Actual times are dependant upon GC column and instrument conditions. (3) Ions (m/z) in bold numbers are suggested primary (quantification) ions. For best signal to noise ratio, do not exceed 10 ions per acquisition group. Dwell times per ion (m/z) is 50 milliseconds. (4) GC peak numbers are those in Figure 1 and Table 10. (5) The list of analytes and internal standards shown is an example. Analyte(s) and internal standard(s) must be selected according to analytical objectives. (6) Retention times are dependant upon GC column and instrument conditions. (7) The better ions for quantification are usually the base peak or those with masses >100 m/z and relative abundances >50% of the base peak. These minimize interference from co-eluting hydrocarbons. The suggested primary ions are not necessarily the base peaks in the mass spectra of the analytes, especially if the base peaks are ions common to aromatics (e.g. m/z 91) and paraffinic or olefinic hydrocarbons (e.g. m/z 42, 57, and 58). Suggested ions for other analytes and internal standards are given in Tables 9 and 10. (8) Secondary ions may be used for quantification if the primary ion encounters interference. Secondary ions improve qualitative identification for SIM analyses. The relative abundances given are approximate (±10 to 20%) and depend upon specific instrument tuning and conditions. They are relative to the primary ion and not necessarily to the base peak in the mass spectrum of each analyte. The relative abundance of secondary ions for each analyte needs to be determined from a mass spectrum acquired on the instrument to be used. (9) (I$) = internal standard. Internal standards must be paired with the appropriate analytes. Tables 8a and 8b give precision and accuracy data for various pairings. Other potentially useful internal standards are given in Table 9. Highly deuterated analogs of the target analytes are preferred, where available. (10) Dibromooctafluorobiphenyl is an optional secondary internal standard useful for monitoring autosampler performance and instrument tuning. A shift in the mass axes or the relative abundance of m/z 296 to that of m/z 456 throughout an analytical sequence will help signal degraded tuning.

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition

Method rev. 1.1.1 METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 16 of 31

Table 7. Suggested spiking schedule for calibration standards and quality control samples Add the following to clean shipping containers (e.g. 50-mL polypropylene centrifuge tubes) in the following order.

Calibration Standards(10) CS0 CS1 CS2 CS3 CS4 CS5 CS6 CS7 CS8 CS9 CS10 Quality Control Samples(11) QB (media blank) QC (matrix spike) QD (matrix spike duplicate)

Volume(2) of Internal Standard Spiking Solution(4,5)

Volume of Target Analyte Spiking Solution(5,6)

Number of Wipes(1,2)

Volume(2) of Isopropanol or Methanol(3)

0 0 0 0 0 0 0 0 0 0 0

3 mL 3 mL 3 mL 3 mL 3 mL 3 mL 3 mL 3 mL 3 mL 3 mL 3 mL

60 µL 60 µL 60 µL 60 µL 60 µL 60 µL 60 µL 60 µL 60 µL 60 µL 60 µL

10 µL 30 µL 100 µL 300 µL 1000 µL

1 1 1

3 mL 3 mL 3 mL

60 µL 60 µL 60 µL

0.0 µL 3-300 µL 3-300 µL

Volume of Spiking Solution diluted 1/20(5,7) 0.0 µL 2 µL 5 µL 10 µL 20 µL 60 µL

or 20-60 µL or 20-60 µL

Volume(2) of Desorption Solution(8)

Resulting µg/sample as Free Base(9)

30 mL 30 mL 30 mL 30 mL 30 mL 30 mL 30 mL 30 mL 30 mL 30 mL 30 mL

0.00 0.02 0.05 0.1 0.2 0.6 2.0 6.0 20 60 200

30 mL 30 mL 30 mL

0.0 0.2-60 0.2-60

(1) Gauze wipes may be added to the calibration standards but are not necessary if cotton gauze is used. Blank gauze wipes must always be added to the quality control samples, QB, QC, and QD. (2) a.) If a sample consists of 2 gauze wipes, the volume of desorption solution must be increased to 40 mL to accommodate the second wipe. The shipping container should be a 50-mL polypropylene centrifuge tube or equivalent to accommodate the extra volume of desorption solution for 2 wipes. It is not critical to know the exact volume of desorption solution and wetting alcohol used per sample. It only needs to be enough to cover the samples and to permit free percolation through the samples. See step 7. b.) If a set of samples consists predominantly of 2 gauze wipes, the QB, QC, and QD should also consist of 2 wipes and treated as per the samples. The volume of isopropanol (or methanol) added to the QC samples should be increased to 4 mL for two gauze wipes to simulate samples containing two gauze wipes. (3) If methanol was used for wipe sampling, it should also be used in the calibration standards, blanks, and QCs instead of isopropanol. (4) Concentration of internal standards in the internal standard spiking solution is approximately 200 µg/mL as the free base. It is critical to know the exact volume of internal standard spiking solution that is added to the calibration standards, samples, blanks, and quality control samples. The volume spiked into the samples may vary with sample size but the volume spiked into each of the calibration standards must not vary. See step 7b. (5) For quality control samples, spike onto wipe media within the shipping container. For liquid calibration standards (in lieu of media calibration standards), spike into the isopropanol (or methanol). (6) Concentration of analytes in the target analyte spiking solution is approximately 200 µg/mL as the free base. (7) Concentration of analytes in the diluted spiking solution for this table is approximately 10.0 µg/mL as the free base and can be prepared by diluting 100 µL of the target analyte spiking solution to 2 mL in methanol. (8) Desorption solution is 0.1 M sulfuric acid in deionized water. (9) This is µg per total sample irrespective of the total desorption solution volume or the area wiped. (10) Select 6 calibration standards from the list to cover the analytical range plus the blank. (11) Prepare one set of quality control samples for every 20 samples or less.

Method rev. 1.1.1

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 17 of 31

Table 8a. Precision and accuracy in scan mode.(1) Compound (D)-Amphetamine

Caffeine

(L)-Ephedrine MDEA MDMA

(D)-Methamphetamine

Phencyclidine

Phentermine (±)-Norephedrine(4) Pseudoephedrine

Internal Standard(2)

Range(3) µg/sample Accuracy

D11-Amp D14-Met NMPhen D11-Amp D14-Met NMPhen D11-Amp D14-Met NMPhen N-PAmp D11-Amp D14-Met NMPhen D11-Amp D14-Met NMPhen D11-Amp D14-Met NMPhen D11-Amp D14-Met NMPhen

0.1-30 0.1-30 0.1-30 1.0-30 1.0-30 1.0-30 0.1-10 0.3-10 0.3-30 0.3-29 0.3-27 0.3-27 0.3-27 0.1-30 0.1-30 0.1-10(5) 0.1-10 0.3-3 0.3-10 0.1-30 0.1-30 0.1-30

D11-Amp D11-Amp D14-Met NMPhen

1-30 0.3-30 0.3-30 0.3-30

17.1 13.7 12.5 20.0 12.5 15.6 15.4 17.8 15.0 16.6 20.2 16.6 22.0 14.7 12.5 14.9 18.2 13.4 16.8 15.2 10.7 9.6 6.5 17.2 14.9 18.7

Bias

Overall Precision

Average

Range

0.0670 0.0610 0.0559 0.0708 0.0636 0.0796 0.0627 0.0674 0.0707 0.0817 0.0778 0.0652 0.0722 0.0631 0.0546 0.0503 0.0690 0.0465 0.0609 0.0486 0.0509 0.0420 0.0328 0.0571 0.0649 0.0488

-0.0613 +0.0338 -0.0310 -0.0832 -0.0014 -0.0040 +0.0510 +0.0666 +0.0293 -0.0224 -0.0739 +0.0589 -0.1017 -0.0435 -0.0348 -0.0665 -0.0683 -0.0577 -0.0682 -0.0720 +0.0190 -0.0269 +0.0061 -0.0783 -0.0422 -0.1068

-0.1048 - -0.0170 -0.0151 - +0.1056 -0.0651 - +0.0177 - 0.1476 - -0.0542 - 0.0274 - +0.0381 -0.0789 - +0.1321 -0.0148 - +0.1128 +0.0261 - +0.1660 -0.0259 - +0.0973 -0.0656 - +0.0657 -0.1011 - -0.0489 -0.0947 - +0.0036 -0.1486 - -0.0315 -0.0657 - -0.0060 -0.1144 - +0.0188 -0.1179 - +0.0110 -0.1257 - -0.0136 -0.0662 - -0.0493 -0.1137 - +0.0091 -0.1010 - +0.0291 -0.0395 - +0.0671 -0.0612 - +0.0340 -0.0070 - +0.0248 -0.1273 - -0.0560 -0.0888 - +0.0395 -0.1505 - -0.0422

(1) Backup Data Report [1]. Values are for chlorodifluoroacetyl derivatives and analysis by GC-MS in scan mode. Each sample consisted of a pair of 3” x 3” 12-ply cotton gauze pads. There were 6 replicate samples per concentration level and six concentration levels evaluated from approximately 0.1 to 30 µg/sample. (2) Internal Standards   Deuterated: Non-deuterated:

D11-Amp = Amphetamine-D11 NMPhen = N-Methyl phenethylamine


D14-Met = Methamphetamine-D14

N-PAmp = N-Propyl amphetamine

(3) Range used for calculation of precision, accuracy, and bias. The entire range studied for all analytes was approximately 0.1 to 30 µg/sample (1xLOQ to 300xLOQ). (4) (±)-Norephedrine = (±)-phenylpropanolamine. (5) One or more higher level concentrations were omitted from the computations due to inlier CVs (<0.0200.)

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition

Method rev. 1.1.1 METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 18 of 31

Table 8b. Precision and accuracy in SIM mode.(1) Compound (D)-Amphetamine

Internal Range(3) µg/ Standard(2) sample Accuracy(4)

Overall(4) Precision

Bias Average

Range

D11-Amp

0.1-30

14.3

0.0412

-0.0750

-0.1153 - -0.0351

D14-Met

0.1-30

10.1

0.0508

-0.0074

-0.0500 - +0.0389

NMPhen

0.1-30

13.3

0.0439

-0.0606

-0.1117 - -0.0318

D11-Amp

0.1-30

21.3

0.0578

-0.1182

-0.1949 - -0.0697

D14-Met

0.1-30

14.4

0.0534

-0.0558

-0.1061 - -0.0170

NMPhen

0.3-30

19.8

0.0387

-0.1338

-0.1775 - -0.0820

D11-Amp

0.3-30

9.1

0.0421

-0.0199

-0.0423 - +0.0157

D14-Met

0.3-30

20.5

0.0503

0.1226

+0.0637 - +0.1883

NMPhen

0.3-30

10.2

0.0449

+0.0260

-0.0075 - +0.0769

MDEA

N-PAmp

0.3-29

10.3

0.0264

-0.0597

-0.0879 - -0.0095

MDMA

D11-Amp

0.1-27

16.2

0.0503

-0.0750

-0.1423 - -0.0292

D14-Met

0.1-0.9(5)

15.4

0.0503(6)

-0.0712

-0.1247 - +0.0032

NMPhen

0.1-27

15.4

0.0496

-0.0722

-0.1136- -0.0108

0.1-10(5)

16.5

0.0379

-0.1030

-0.1414 - -0.0660

D14-Met

0.1-30

9.2

0.0351

-0.0343

-0.0767 - +0.0006

NMPhen

0.1-30

13.6

0.0322

-0.0827

-0.1221 - -0.0403

D11-Amp

0.1-10

17.7

0.0428

-0.1068

-0.1303 - -0.0586

D14-Met

0.1-3

11.3

0.0450

-0.0393

-0.0683 - -0.0205

NMPhen

0.1-3(5)

16.1

0.0449

-0.0871

-0.1279 - -0.0383

D11-Amp

0.1-30

12.8

0.0394

-0.0637

-0.0982 - -0.0433

D14-Met

0.1-30

9.8

0.0495

-0.0051

-0.0375 - +0.0556

NMPhen

0.1-30

11.0

0.0394

-0.0451

-0.0766 - -0.0163

D11-Amp

0.3-30

17.3

0.0402

-0.1073

-0.1496 - -0.0514

D14-Met

0.3-30

11.7

0.0519

-0.0294

-0.0559 - +0.0532

NMPhen

0.3-30

17.0

0.0450

-0.0956

-0.1197 - -0.0576

Caffeine

(L)-Ephedrine

(D)-Methamphetamine D11-Amp

Phencyclidine

Phentermine

Pseudoephedrine

(5)

(1) Backup Data Report [1]. Values are for chlorodifluoroacetyl derivatives and analysis by GC-MS in SIM mode (see Table 6 for MS conditions). Each sample consisted of a pair of 3” × 3” (7.5 cm x 7.5 cm) 12-ply cotton gauze pads. There were 6 replicate samples per concentration level and six concentration levels evaluated from approximately 0.1 to 30 μg/sample. Norephedrine (phenylpropanolamine) was not evaluated in the SIM mode due to breakdown at room temperature storage for several days prior to analysis. (2) Internal Standards, Deuterated: Non-deuterated: D11-Amp = Amphetamine-D11 NMPhen = N-Methyl phenethylamine

D14-Met = Methamphetamine-D14 N-PAmp = N-Propyl amphetamine


(3) Range used for calculation of precision, accuracy, and bias. The entire range studied for all analytes was approximately 0.1 to 30 μg/sample (1×LOQ to 300×LOQ). (4) NIOSH [1995]. NIOSH Technical Report: Guidelines for Air Sampling and Analytical Method Development and Evaluation. By Kennedy ER, Fischbach TJ, Song R, Eller PM, Shulman SA. Cincinnati, OH: U.S. Department of Health and Human Services, Public Health Services, Centers for Disease Control and Prevention, National Institute for Occupational Safety and Health, DHHS (NIOSH) Publication No. 95-117. (5) One or more higher level concentrations were omitted from the computations due to inlier CVs (<0.0200). (6) The overall precision, ŜrT, is an estimate due to inlier precisions (<0.02) at several higher concentration levels.

Method rev. 1.1.1

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 19 of 31

Table 9a. Recommended internal standards and best application(1) COMPOUND NAME

CAS

(±)-Amphetamine-D11 (±)-Amphetamine-D8 (±)-Amphetamine-D6 (±)-Methamphetamine-D14 (±)-Methamphetamine-D11 (±)-Methamphetamine-D9 N-Methylphenethylamine Phencyclidine-D5 MDEA-D6(2) N-Propylamphetamine(2)

Not available 145225-00-9 Not available Not available 152477-88-8 Not available 589-08-2 60124-86-9 160227-44-1 Not available

MW as free base 146.12 143.15 141.16 163.12 160.15 158.16 135.23 248.35 213.22 177.29

Quant. Ion

Secondary Ion

COMMENTS

128 159(3) 123 128 126 123 104 247 165 156

Preferred analog for amphetamine Alternate for amphetamine-D11 Alternate for amphetamine-D11 Preferred methamphetamine analog Alternate for methamphetamine-D14 Alternate for methamphetamine-D14 Alternate for methamphetamine-D14 Use only for phencyclidine. Use only for MDEA. Alternate for MDEA-D6

160 126(3) 160 177 176 177 156 205 190 198

(1) Care must be exercised in the selection of internal standards for each analyte because of differences in derivatization efficiencies due to structural differences. a. Deuterated analogs of each target analyte may be acceptable as internal standards if they are isotopically pure enough and their ions do not interfere with the quantification ions (usually base peaks) of the target analyte, especially at the limit of detection for the target analyte. Conversely it is also important that ions in the target analyte, especially at high concentrations, do not interfere with the quantification ion (usually base peaks) of any deuterated analog used as the internal standard. b. The more highly deuterated an analog, the more it will chromatographically separate from the target analyte, reducing interference from common ions. c. Phentermine and mephentermine have been used as internal standards. Such use is not advised in this method because of their reported occasional use as adulterants in certain illicit drugs such as MDMA. (2) N-Propylamphetamine and MDEA-D6 are only applicable to MDEA and other hindered amines (e.g. fenfluramine and MBDB) due to similar steric hindrance at the nitrogen (N-ethyl or N-propyl substitution) which affects derivatization efficiency. (3) It is better to use m/z 126 because at high concentration levels unlabelled amphetamine contributes significant interference to m/z 159 of amphetamine-D8.

Table 9b. Recommended best application of internal standards Recommended Deuterated Internal Standards

TARGET ANALYTE Amphetamine Caffeine Ephedrine MDEA MDMA Methamphetamine Phencyclidine Phentermine (±)-Norephedrine(4) Pseudoephedrine

Amphetamine-D11(2) X X

Methamphetamine-D14(2) X X X

MDEA-D6(1)

Recommended Alternate Non-deuterated Internal Standards(3) Phencyclidine-D5

N-Methylphenethylamine X X X

4-Phenyl-1butyl-amine

X

X X

X X X

X X

N-Propyl-amphetamine(1)

X

X

X X X X X

X

X

(1) N-Propylamphetamine and MDEA-D6 are only applicable to MDEA and other hindered amines (e.g. fenfluramine and MBDB) due to similar steric hindrance at the nitrogen (N-ethyl or N-propyl substitution) which affects derivatization efficiency. (2) The alternate deuterated compounds listed in part A above may be used. Avoid ring-labeled amphetamine-D5 (CAS 65538-33-2) since the primary (quantification) ion is the same as for amphetamine and GC peaks overlap significantly. Also avoid methamphetamine-D5 (CAS 60124-88-1) since GC peaks significantly overlap and secondary ions for the chlorodifluoroacetyl derivative are not baseline resolved. (3) The listed non-deuterated compounds are effective as internal standards for the listed target analytes. Non-deuterated internal standards might not be permissible. (4) (±)-Norephedrine is the same as (±)-phenylpropanolamine.

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition

Method rev. 1.1.1 METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 20 of 31

Table 10. Gas chromatographic retention times for chlorodifluoroacetyl derivatives of amphetamines, precursors, adulterants, and miscellaneous drugs of abuse.(1) GC Peak No.(2) 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41

Compound Nicotine (DL)-Amphetamine-D11(I$)(7) (DL)-Amphetamine Phenethylamine(8) Phentermine(8) N-Methyl pseudoephedrine(9) N-Methyl pseudoephedrine(9) N-Methyl phenethylamine (I$)(7) (DL)-Methamphetamine-D14 (I$)(7) (DL)-Methamphetamine Fenfluramine(8) S-(-)-Cathinone (from Khat plant) Bupropion (Wellbutrin®, Zyban®) N-Ethyl amphetamine Ecgonine, methyl ester S-(-)-Methcathinone (“Cat”) Norpseudoephedrine (Cathine) (±)-Norephedrine Aminorex Dibromooctafluorobiphenyl (I$)(7) N-Propyl amphetamine (I$)(7) 4-Methoxyamphetamine 4-Phenyl-1-butylamine (I$)(7) 1S,2R(+)-Ephedrine-D3 (I$)(7) (DL)-Ephedrine Acetaminophen(8) Methyl phenidate (Ritalin®) Pseudoephedrine Meperidine (Demerol® etc.) Atropine (±)-MDA Caffeine(8) N,N-Dimethyltryptamine (DMT) (±)-BDB Ketamine (“special K”)(8,11) Lidocaine(8) Trifluoromethylphenyl piperazine(11) Benzyl piperazine(11) (“Legal XTC”) Phencyclidine-D5 (I$)(7) Phencyclidine (PCP) MDMA(11)

Recommended Quantification (1’) and Confirmation(2’, 3’) Ions (m/z)(3) 1’ 2’ 3’ 84 133 162 160 128 162 156 118 158 104 91 170 172 132 134 162 75 72 156 104 158 177 98 179 170 172 118 184 186 159 105 77 132 44 100 111 184 186 118 182 82 311 170 105 172 156 158 246 156 158 246 107 79 232 296 456 454 198 156 200 121 148 78 176 104 173 175 85 170 172 260 108 221 263 84 56 91 170 172 260 71 247 172 124 94 103 135 162 291 194 109 67 58 129 102 135 176 170 180 182 209 86 58 120 200 145 172 91 197 175 205 96 246 200 242 243 170 162 135

Retention Relative Time in Retention Form(4) Minutes Time(5) parent 8.92 0.396 derivative 10.26 0.800 derivative 10.34 0.807 derivative 10.38 0.810 derivative 10.52 0.821 derivative 10.54 0.822 parent abt 11 0.86 derivative 11.37 0.887 derivative 11.70 0.913 derivative 11.79 0.920 derivative 11.83 0.923 derivative 11.99 0.935 parent 12.14 0.947 derivative 12.22 0.953 derivative 12.36 0.964 derivative 12.38 0.966 bis-derivative 12.46 0.972 bis-derivative 12.49 0.974 derivative (-CN) 12.70 0.991 parent 12.82 1.000 derivative 12.97 1.012 derivative 13.22 1.031 derivative 13.27 1.035 derivative 13.44 1.048 bis-derivative 13.48 1.052 derivative 13.67 1.066 parent 13.81 1.077 bis-derivative 13.93 1.087 parent 13.99 1.091 parent (-H2O) 14.25 1.112 derivative 14.36 1.120 parent 14.84 1.158 derivative 14.97 1.168 derivative 15.11 1.179 parent 15.20 1.186 parent 15.28 1.192 derivative 15.46 1.206 derivative 15.54 1.202 parent 15.59 1.216 parent 15.62 1.218 derivative 15.66 1.221

Relative Retention Time(6) 0.757 0.870 0.877 0.880 0.892 0.894 0.93 0.964 0.992 1.000 1.003 1.017 1.030 1.036 1.048 1.050 1.057 1.059 1.077 1.087 1.100 1.121 1.126 1.140 1.143 1.159 1.171 1.182 1.187 1.209 1.218 1.259 1.270 1.282 1.289 1.296 1.318 1.318 1.322 1.325 1.328

Table 10 continued,

Method rev. 1.1.1

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 21 of 31

Table 10, continued. Gas chromatographic retention times for chlorodifluoroacetyl derivatives of amphetamines, precursors, adulterants, and miscellaneous drugs of abuse.(1) GC Peak No. (2) 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67

Compound MDEA-D6 (I$)(7) MDEA(11) Phenylephrine(8) (±)-MBDB Theophylline(8) Mescaline Phenylephrine(8) Chlorpheniramine(8) Methyl phenidate 4-Bromo-2,5-DMPEA(10) (Nexus) cis-(±)-4-Methylaminorex (“U4Euh”) Dextromethorphan(8) Methaqualone Cocaine Atropine(8) Diazepam (Valium® etc.) Hydrocodone (Lortab® etc.) Hydromorphone (Dilaudid®) Hydrocodone (Lortab® etc.) Morphine Codeine Oxycodone (OxyContin®) Hydromorphone (Dilaudid®) Flunitrazepam (Rohypnol®, roofies)(11) Morphine Fentanyl (Sublimaze® etc.)

Recommended Quantification (1’) and Confirmation (2’, 3’) Ions (m/z)(3) 1’ 2’ 3’ 190 165 135 184 162 135 156 158 374 184 176 135 180 95 68 181 194 179 156 248 158 203 205 167 196 198 242 244 229 203 160 117 271 59 150 235 250 233 82 182 303 124 82 94 256 283 284 299 242 284 285 228 229 411 354 298 268 397 269 282 411 283 315 230 316 397 341 398 312 285 286 380 382 509 245 146 189

Form(4) derivative derivative tris-derivative derivative parent derivative bis-derivative parent derivative derivative derivative parent parent parent derivative parent parent parent derivative derivative derivative parent derivative parent bis-derivative parent

Retention Time Minutes 16.01 16.04 16.10 16.29 16.34 16.43 16.65 16.73 17.20 17.57 17.89 18.09 18.27 18.62 19.10 20.76 20.91 21.04 21.13 21.20 21.28 21.57 21.78 22.19 22.26 22.96

Relative Retention Time(5) 1.249 1.251 1.256 1.271 1.275 1.282 1.299 1.305 1.322 1.370 1.396 1.411 1.425 1.452 1.490 1.619 1.631 1.641 1.648 1.654 1.660 1.682 1.699 1.731 1.736 1.791

Relative Retention Time(6) 1.358 1.360 1.366 1.382 1.386 1.394 1.412 1.419 1.459 1.490 1.517 1.534 1.550 1.579 1.620 1.761 1.774 1.785 1.792 1.798 1.805 1.830 1.847 1.882 1.888 1.947

(1) Actual retention times may vary depending on individual GC column and GC conditions. Gas chromatographic conditions used are on p. 9106-1. The mass spectrometer was operated under the conditions given on p 9106-1 (or see the Backup Data Report [1].) (2) GC peak numbers represent peaks as numbered in Figure 1. (3) Use extracted ion chromatograms of the primary ions (1’) for quantifying peaks in either the scan mode or the SIM mode. Use the secondary and tertiary ions (2’ and 3’) for qualitative identification when necessary. These ions are selected for nearness to the primary ion to minimize false negatives from skewed spectra and from low mass interference from hydrocarbons. (4) Not all forms are presented. Parent compounds are not presented that have irregular or overly broad GC peak shapes under the GC conditions used. Spectra for chlorodifluoroacetyl derivatives are given in the Backup Data Report [1] (5) Retention time relative to 4,4’-dibromooctafluorobiphenyl. (6) Retention time relative to the chlorodifluoroacetyl derivative of methamphetamine. (7) I$ = Internal standard. (8) Intentional or unintentional adulterants. For example, phentermine may be added to MDMA and caffeine added to methamphetamine. Chlorpheniramine is an unintentional adulterant when pseudoephedrine containing chlorpheniramine is used as a methamphetamine precursor. (9) Presence of (+)-Norephedrine, N-methylpseudoephedrine and/or N-methylephedrine in pseudoephedrine or ephedrine indicates extracts of Ephedra species (spp.) as source. Presence of amphetamine and N,N-dimethylamphetamine in methamphetamine final product also indicates the same source.[18, 19, 20] (10) 4-Bromo-2,5-dimethoxyphenethylamine (11) Typical “club drugs” (piperazine analogs as ecstasy substitutes, ketamine and flunitrazepan as predatory drugs).

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition

Method rev. 1.1.1 METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 22 of 31

Table 11a. Recovery from latex-painted wall with various solvents; one gauze wipe compared with the sum of two gauze wipes(1,2) Water(3) First Gauze Wipe

Test Compound(5)

Percent

%RSD

Amphetamine

51

14

Cocaine

36

Ephedrine MDMA

Isopropanol Plus Second Wipe(4) Percent

Methanol

Plus Second Wipe(4)

First Gauze Wipe

Percent

First Gauze Wipe

Percent

%RSD

Plus Second Wipe(4)

Percent

%RSD

Percent

56

67

6.0

78

90

4.0

96

22

36

69

22

80

89

9.1

94

48

23

52

76

7.4

85

91

4.4

96

40

20

44

61

9.0

70

88

5.3

94

MDEA

45

22

50

69

12

80

90

11

97

Methamphetamine

46

16

50

64

7.4

75

87

3.5

94

Phencyclidine

27

26

30

64

9.6

73

86

5.2

91

Phentermine

53

9.2

58

78

6.6

91

95

2.9

101

Phenylpropanolamine

58

21

62

80

9.3

95

85

5.0

94

Pseudoephedrine

49

20

53

73

7.0

85

95

3.3

101

(1) Backup Data Report for NIOSH 9109 [8]. Area of each sample was 100 cm . (2) Wall was an existing standard gypsum board wall painted with a latex based paint. Painted surface was at least one year old. There were six replicates for each solvent tested. (3) Water was deionized water (ASTM type II). Note low recovery and high %RSD. (4) For the serial wipe study, each 100-cm2 area was wiped again with a fresh pre-wetted gauze wipe and the amount recovered was determined separately. In practice, a second (serial) wipe is included with the first gauze wipe; both gauze wipes constitute a single sample. The percent recoveries shown in the column represent the sum of the amounts recovered in both the first and second wipes. (5) Each pre-measured area was spiked with 3 µg of each analyte in methanol and the methanol allowed to dry for several minutes prior to wipe sampling. 2

Table 11b. Recovery from various surfaces with various solvents; one gauze wipe compared with the sum of two gauze wipes(1) Isopropanol First Gauze Wipe

Surface Material(3) Enamel (lid of washing machine) Vinyl veneer on particle board Latex painted wall Refrigerator door Varnished hardwood panel Formica® countertop

Replicates 4(3) 4(4) 6(3) 2(4) 2(5) 4(4)

Percent 58 60 64 65 72 75

%RSD 5.7 5.2 7.4 2.9 5.4 4.9

Methanol

Plus Second Wipe Percent 68 68 75 76 76 82

(2)

First Gauze Wipe

Percent 81 81 87 91 82 87

%RSD 2.4 4.8 3.5 4.0 3.7 3.8

Plus Second Wipe(2) Percent 87 89 94 92 86 91

(1) Backup Data Report for NIOSH 9109 [8]. Area of each sample was 100 cm2. (2) For the serial wipe study, each 100-cm2 area was wiped again with a fresh pre-wetted gauze wipe and the amount recovered was determined separately. In practice, a second (serial) wipe is included with the first gauze wipe; both gauze wipes constitute a single sample. The percent recoveries shown in the column represent the sum of the amounts recovered in both the first and second wipes. (3) The Refrigerator door and the washing machine lid were from used appliances. The vinyl-veneered particle board (a book shelf), the Formica® countertop, and the varnished hardwood paneling were all purchased new. All surfaces of used and new materials were pre-cleaned with multiple rinses of methanol prior to spiking. Each pre-measured 100-cm2 square was spiked with 3 µg methamphetamine. (4) Samples were taken using the side-to-side and then top-to-bottom wiping technique. (5) Half of the samples were wiped using the side-to-side wiping technique and half were wiped using the concentric squares wiping technique. There were no significant differences in recoveries. Percent recoveries and %RSDs are for both techniques combined. (6) Samples were taken each time using only top-to-bottom wiping with the grain of the wood in an “N” pattern. Method rev. 1.1.1

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 23 of 31

Figure 1. Typical chromatograms of chlorodifluoroacetyl derivatives by GC-MS in scan mode

Figure 1a. Extracted ion chromatogram for m/z 156 (155.70 to 156.70). Figure 1b. Extracted ion chromatogram for m/z 170 (169.70 to 170.70). Figure 1c. Total ion chromatogram (TIC). GC Peak Identification: See Table 10 for identification of numbered GC peaks. (But note that retention times in Table 10 do not correspond to those in Figure 1 because a different DB-5 column was used.) GC-MS Conditions: See p. 9106-1 for GC-MS conditions.

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition

Method rev. 1.1.1 METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 24 of 31

APPENDIX A. REAGENTS and SOLUTIONS: 1. For derivatization, pentafluoropropionic anhydride (PFPA) may be substituted for chlorodifluoroacetic anhydride (CDFAA). Spectra, retention times, suggested quantification ions, and precision and accuracy data for PFPA derivatives are given in the Backup Data Report [1]. Spectra for CDFAA derivatives are also given in the Backup Data Report [1]. NOTE: 100 μL of pentafluoropropionic anhydride (PFPA) may be substituted for chlorodifluoroacetic anhydride, but the samples must be heated to 90 °C for 20-30 minutes in step 11c. 2. The instrumental internal standard, 4,4’-dibromooctafluorobiphenyl is optional. It is useful for monitoring instrument tuning and autosampler performance. 3. Primary amines form Schiff bases and enamines with ketones and aldehydes. These may in turn form derivatives with the acylating reagents. The use of acetone must be avoided strictly prior to the analytes being derivatized. Glassware and equipment rinsed with acetone must be thoroughly dried. Toluene should be avoided for making up standard solutions because it usually contains benzaldehyde, an oxidation product of toluene. Condensation products have been observed between primary amines and benzaldehyde. The only solvents recommended for the preparation of stock solutions and dilutions thereof are methanol (preferably) and isopropanol. 4. The reconstitution solvent should not contain methanol or other alcohol since the derivatized alcohol groups in ephedrine type compounds are hydrolyzed over time. Toluene containing 10 percent acetone is recommended. B. EQUIPMENT: 1. Wipe media: Besides cotton gauze, 4”×4” (10 cm x 10 cm) 4-ply MIRASORB® (Johnson and Johnson), and 4”×4” (10 cm x 10 cm) AlphaWipe® (TX® 1004, Texwipe Corp.) were acceptable wipe media and can be used in the absence of cotton gauze. MIRASORB®, a non-woven cotton/polyester blend, is discontinued but counterparts exist that claim to be of identical construction and fiber composition. AlphaWipe® is a hydrophilic, highly adsorbent, tightly knitted continuous filament polyester wipe. Precision and accuracy data for MIRASORB® and AlphaWipe® are given in the Backup Data Report [1]. 2. Shipping containers: The 50-mL polypropylene centrifuge tubes with caps are preferred for one or two gauze wipes and are not as breakable as glass 40-mL VOA vials. The 40-mL VOA vials are acceptable for single gauze wipes. Larger containers (glass with a PTFE lined cap) should be used for combining more than two gauze wipes into a single sample. The size of the container for two or more wipes should be approximately 25 mL per gauze wipe (e.g. a minimum size of 100-mL for up to four gauze wipe samples). There needs to be enough extra headspace in the shipping container to allow the desorption solution to cover the gauze wipes and to percolate freely through the wipe sample(s) during mixing. 3. Each regulatory agency having legal jurisdiction over the contaminated site may require different but specific off-site preparation and on-site sampling procedures. It is important to consult local regulatory agencies or departments of health having legal jurisdiction over contaminated sites to determine specific sampling, quality control, analyses, and reporting requirements. C. SAMPLING: 1. Follow specific requirements of surface area to be wiped (usually 100 cm2 or 1000 cm2) and action threshold (or maximum allowable residual level) set by the state or specified by the client. Uptake rates depend upon the wipe sampling method used, so the specific wipe technique used must be specified, and any deviations from the required wipe sampling requirements noted. NOTE: To ensure that samples have not been tampered with, the use of custody seals and a chain-ofcustody form is strongly recommended. 2. Prepare a rigid template from disposable cardstock or a sheet of PTFE having either a 10 cm × 10 cm or 32 cm × 32 cm square-cut hole. The template must be able to retain its shape during wiping to ensure that the areas wiped were either 100 cm2 or 1000 cm2. Secure the template(s) to the area(s) to be wiped (e.g. with tape along outside edge of template). If a single-use disposable template is not used, clean the template between samples to avoid cross-contamination, and provide the Method rev. 1.1.1

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 25 of 31

laboratory with a blank wipe of the cleaned template between samples to determine that no crosscontamination has occurred. 3. A template might not always be applicable, as in curved or odd-shaped areas such as around burners on stove tops or a fan blade. In such cases sample an area as close to either 100 cm2 or 1000 cm2 as feasible and provide the measurement to the regulatory agency and to the analytical laboratory for proper reporting. Tape can be used to delineate the sampling area. 4. It is recommended to provide extra wipe media from the same lot for required media blanks, field equipment blanks, samples, and quality-control samples. 5. Gauze in sterile packaging is recommended to minimize the chance for cross-contamination, which can more easily occur with open bulk packaged cotton gauze. 6. To prevent contamination in the field, another alternative is to pre-wet and insert the gauze wipes into the sample containers off-site. This avoids any possibility of the bottle of methanol or isopropanol becoming contaminated on-site with methamphetamine (or other analytes). If the wipes were prepared off-site, then remove pre-wetted gauze wipe from sample container, opening only one sample container at a time. In either case, squeeze out and discard any excess solvent from the gauze wipe. Use fresh latex or nitrile gloves for each separate sample and blank. Do not use vinyl gloves due to the potential for leaching of phthalate plasticizers and contamination of the samples. 7. Wipe techniques a. Concentric Squares Wiping Technique (particularly suitable for smooth and non-porous surfaces): Fold the pre-wetted gauze in half and then fold in half again. Using firm pressure wipe the area within the template. Start at one of the inside corners of the template and wipe in concentric squares, progressing toward the center. End with a scooping motion. Without allowing the gauze to touch any other surface, reverse the last fold so that the exposed side of the gauze is facing inward and using a fresh surface of the gauze, wipe the same area in the same manner as before. Roll or fold the gauze again and insert into the shipping container. Note: Wiping in concentric squares is described by OSHA [21]. It is especially suitable for large (e.g. 1000 cm2) areas. b. Side-to-side Wiping (or Blotting) Technique (particularly suitable for rough, porous, and/or soiled surfaces): Fold the pre-wetted gauze in half and then fold in half again. Using firm pressure wipe or blot the area within the template with at least five overlapping side-to-side horizontal passes (see NOTE) beginning at the top and progressing to the bottom in a “Z” pattern. End with a scooping motion. If blotting, blot at least five times on each horizontal pass (see NOTE). Without allowing the gauze to touch any other surface, reverse the last fold so that the exposed side of the gauze is facing inward. Using a fresh surface of the gauze, wipe or blot the area again with at least five overlapping top-to-bottom vertical passes beginning at the left side and progressing to the right in an “N” pattern. If blotting, blot at least five times on each vertical pass. Roll or fold the gauze again and insert into the shipping container. Blotting is suggested in areas so soiled or rough that the threads of the gauze media are continually snagged. NOTE: On areas larger than 100 cm2, more than five passes and blots will be needed. c. Repeat or Serial Wiping: If isopropanol is used for wiping, a serial or repeat wipe sample of the same area with a fresh gauze wipe will improve sampling efficiency. (See recoveries for second wipe in Tables 11a and 11b.) For serial wiping, repeat the wiping procedure described above (APPENDIX steps 7a or 7b) with a fresh gauze wipe. Place the second gauze wipe into the same shipping container as the first gauze. The 50-mL polypropylene centrifuge tubes are large enough to contain up to two gauze wipes. NOTE: If the area to be wiped remains substantially wet from the first gauze, the second gauze wipe might be used in the dry state to soak up the residual solvent from the first gauze wipe. 8. Composite sampling: Composite samples are allowed by some regulatory agencies. Their use for quantitative purposes may be subject to the permission and guidance of regulatory agencies. Refer to guidelines of regulatory agency for directions on composite sampling. A basic default guideline for composite sampling is as follows: Do not mix inconsistent samples; that is, areas wiped must NIOSH Manual of Analytical Methods (NMAM), Fifth Edition

Method rev. 1.1.1 METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 26 of 31

be equal in area, sampled areas must have the same high or low probability of contamination, and sampled areas must relate to a specific target appliance or site and not to several appliances or incongruous sites combined. NOTE: Composite samples cannot meet specific action-threshold requirements for discrete sampling locations. Nor do composite samples consisting of four wipes, for example, improve the sensitivity by decreasing the LOD four fold; instead it raises the LOD by a factor related to the extra volume of desorption solution that is required to desorb a larger number of wipes. The following example illustrates these two points. Assume that the action level was 0.1 μg/100 cm2. If the analysis gave an LOD of 0.06 μg/sample for a single wipe or discrete sample covering an area of 100 cm2, then the LOD for the analysis could be expressed as 0.06 μg/100 cm2, which is low enough to be able to determine whether any discrete sample is at or exceeds the action level. Now if a composite of four wipes was taken, each with an area of 100 cm2 for a total area wiped of 400 cm2, the LOD for that composite sample is not 0.06 μg/400 cm2 nor is it 0.015 μg/100 cm2; it is actually several times larger than 0.06 μg/400 cm2. First of all it increases relative to the ratio of the volume of desorption solution used to desorb the sample compared to that used for the calibration standards. Secondly it has nothing to do with the area that was wiped, because the LOD for the calibration curve is determined in terms of μg per sample, independent of the area. To explain the first point, assume approximately 90 mL was used (for ease in calculation) to desorb the four wipes and 30 mL (the normal amount for a single wipe) was used to desorb each calibration standard. The calculation of the LOD for the four composited samples would be μg/sample × (desorption volume for 4 wipes) ∕ (desorption volume for the calibration standards), or 0.06 μg/sample × (90 mL/30 mL), or 0.18 μg/sample for the composited sample. Since the area wiped for the composite sample was 400 cm2, the LOD for that sample could be expressed as 0.18 μg/400 cm2. Regarding the second point, this value, 0.18 μg/400 cm2, cannot be construed or mathematically reduced to 0.045 μg/100 cm2 because it cannot be known whether three of the four wipes were blank and the fourth wipe just under the value of 0.18 μg. Hence, the effective LOD per individual wipe has to be regarded not only as 0.18 μg/400 cm2 but also as 0.18 μg/100 cm2 because any value determined for entire 400 cm2 might have come from just one of those 100 cm2 areas. Thus, for composite samples, the LOD must be expressed in terms of the entire area wiped and not extrapolated to some portion thereof. In this example, an LOD of 0.18 μg/100 cm2 is above the action threshold of 0.1 μg/100 cm2, meaning that this composite sample cannot satisfy the requirement that residual levels be below 0.1 μg/100 cm2. It remains for the regulatory agency and not the laboratory to determine how to apply results for composite samples to the established action levels. The same consideration that is given above for the LOD applies to results that are greater than the LOD. To avoid confusion in reporting concentrations for composite samples, it is recommended that the sample concentration (in μg/sample, whatever the sample size) and the total area wiped (in cm2) be reported separately. For example, a result of 0.4 μg/sample for a sample consisting of four separate wipes of 100 cm2 each (for a total area wiped of 400 cm2), is to be reported as 0.4 μg/400 cm2 and not averaged to 0.1 μg/100 cm2. This manner of reporting may be required by some regulatory agencies. 9. For quality assurance purposes, regulatory agencies may require duplicate samples to be taken in the field. If such is the case, an area contiguous with and adjacent to the first area, if possible, should be wiped as described under SAMPLING. Do not re-wipe the previously wiped area. This sample is a blind sample and should not be identifiable by the analytical laboratory as a duplicate of any other sample. These are distinct from the laboratory duplicates of a single sample described in step 14 of the method. Field duplicates are useful for evaluating the consistency of sampling technique, assuming uniformity of contamination on adjacent sampling sites. Laboratory duplicates are useful for evaluating consistency of sample preparation and instrumental analysis.

Method rev. 1.1.1

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 27 of 31

D. DESORPTION FROM MEDIA: 1. An internal standard spiking solution volume of 60 μL was selected for ease in scaling from 60 μL per 30 mL to 80 μL per 40 mL of desorption solution. In either case the rate of 2 μL internal standard spiking solution per mL desorption solution was used. However, any convenient volume of internal standard spiking solution (e.g. 50 μL) that can be delivered reproducibly is acceptable. Whatever volume is chosen, there must be no variation in the volume of the internal standard spiking solution used in preparing each of the calibration standards. If spiking Strategy A is used (see APPENDIX D3), it is critical to know the exact volume of internal standard spiking solution that is applied to each sample (V1), the media blanks (V5), and the calibration standards (V2), since these volumes are used for internal standard spiking solution volume corrections in step 19. 2. It is not necessary to know the exact volume of desorption solution added to each sample or the volume of residual wetting alcohol because differences in the volumes are normalized through the use of internal standards added prior to desorption. 3. Alternate strategy for spiking internal standards (spiking strategy B below): By using the exact same volume of internal standard spiking solution in all samples, blanks, QC samples, and calibration standards, regardless of the volume of desorption solution added or residual wetting alcohol, the volume corrections in step 19, (V1/V2 and V5/V2) drop out of the equation. However, the internal standard GC peak areas must still be measurable in samples where larger volumes of desorption solution are used (such as for composite samples). Because of the increased dilution of the internal standard in larger samples, this approach should be limited to desorption solution volumes of about 120 mL or less. NOTE: There are two separate strategies for handling larger samples requiring larger volumes of desorption solvent. These are outlined below as strategies A and B. Volume of Internal Standard Spiking Solution (μL)

Number of Wipes

Size of Shipping Container (mL)

Strategy A

Strategy B

Volume of Desorption Solution (mL) (Strategies A and B)

1 2 4 (e.g., Composite)

40-50 50 100-120

60 80 160

60 60 60

30 40 80

Apply volume correction factors at step 19.

Do not apply volume correction factors at step 19.

With either strategy, if two gauze wipes were included in the samples, then use 40 mL of desorption solution. If four gauze wipes were included in the samples, then use 80 mL of desorption solution. a. In strategy A, the volume of internal standard spiking solution is kept at a constant ratio of 2 μL per mL of desorption solution added. This enables larger samples to be desorbed without diminishing the area of the GC peak for the internal standard. However, a volume correction factor (V1/V2) is needed in the final calculations in step 19. Therefore, the exact volume of internal standard added to each of the samples relative to that added to the calibration standards must be known. b. In strategy B, the volume of internal standard spiking solution is kept constant for all samples and calibration standards, but need not be exactly 60 μL. This enables the final calculations to be made in step 19 without a volume correction factor. However, the area of the GC peak for the

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition

Method rev. 1.1.1 METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 28 of 31

internal standard will vary with sample desorption volume and the internal standard must be concentrated enough to be measurable where larger volumes of desorption solution are used. E. DRYING COLUMN PREPARATION Using 1 cm i.d. × 12-15 cm long polypropylene columns having a fritted polyethylene disc or equivalent (see EQUIPMENT), add 1 gram (~0.8 cc) of anhydrous potassium carbonate (the bed dimension will be about 1.0 cm dia. × 1 cm long). Add 1 gram (~0.8 cc) anhydrous sodium sulfate on top of the potassium carbonate. Remove any particles clinging electrostatically to the outside surfaces. NOTE: Particles of the drying salts must not get into the collection tube, either through the frits or glass wool plugs, or from particles clinging electrostatically to the outside of the columns. Salts appear to inhibit derivatization efficiency. F. DERIVATIZATION: If isopropanol was used as the wetting solvent for the wipes, some of it will be co-extracted into the methylene chloride. In the presence of trace isopropanol, the crystal violet will go through a series of color changes as the extracts are evaporated to dryness. However, if methanol was used as the wetting solvent, the color of the crystal violet will remain blue to blue-violet at all stages of drying. Yet even with methanol, the same color changes can be afforded by adding 0.1 mL of isopropanol to the extracts prior to evaporation. Recoveries of analyte will not be affected in the absence of isopropanol, however, as long as the residues are dry before proceeding to step 11b. With the presence of a small amount of added or co-extracted isopropanol, as each sample concentrates, the color of the solution will go from a blue or violet color rapidly through green to a yellow color as the residue approaches dryness, which is indicative of increasing hydrogen ion concentration in the residual alcohol. Upon continued blowing with nitrogen, the color of the residue turns back to a green or blue hue just at the point of dryness, which is indicative of the loss of excess hydrogen chloride and/or alcohol. At this stage the samples are dry and may be removed. Continued blowing beyond this point may turn the dried residue to a deep blue-violet or violet color. Losses of analyte have not been experienced even after blowing for five additional minutes beyond the violet stage as long as the hydrochloric acid had been added. Color changes will not be as dramatic or will not develop if too much crystal violet is used. As the samples become concentrated, the tubes may be raised up in the water bath so that only the very bottoms of the tubes touch the surface of the water. This makes it easier to observe the color changes. The tubes may be raised out of the water bath, but blow-down times are lengthened. Prolonged heating at high temperatures during derivatization with the acidic conditions of the acid anhydride derivatizing agents promotes mutual isomerization between the ephedrine diastereomers (ephedrine and pseudoephedrine). Dehydration of the ephedrine compounds (ephedrine, norephedrine, and pseudoephedrine) also occurs to some extent to yield β-amino-β-methyl styrenes. Heating during derivatization for longer than one hour is especially not recommended. Thirty minutes is sufficient. NOTE: The color of the solution will gradually fade from purple to deep blue within about 20-30 minutes. This is due to the known tendency of phenolphthalein to fade at high pH. It has also been observed that in certain bulk samples, unknown constituents will cause the color of phenolphthalein to fade rapidly so that a purple color cannot be obtained at a pH >9, leaving only the blue color of the bromothymol blue. A quick check with pH paper can confirm that the pH is 9 or greater. G. MEASUREMENT: Recoveries for the laboratory control matrix spike samples (QC and QD) must meet the guidelines of the specific regulatory agency involved, if applicable (80-120% is a reasonable target in the absence of specific guidance).

Method rev. 1.1.1

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 29 of 31

NOTE: The QC samples (QC and QD) in this method may be referred to in some guidance documents as matrix spike and matrix spike duplicate samples (MS/MSD), but serve the same purpose. Analyze and report field-equipment blanks as samples. Do not subtract their values from any other sample. Recoveries of Continuing Calibration Verification (CCV) standards must meet guidelines of regulatory agency (80-120% is a reasonable target in the absence of specific guidance). The CCV standards may be referred to in some guidance documents as “QC samples,” but such “QCs” are equivalent to liquid standards (not matrix spiked samples) and serve the same purpose of the CCVs in this method. With the GC/MS it is possible to achieve the lower limit of 0.05 μg or less per sample for methamphetamine in either the scan mode or SIM mode. The scan mode is essential where the identification of unknowns is an analytical objective. If lower limits of detection are desired or difficult to obtain in the scan mode, or for routine target compound only analyses, the instrument may be operated in the SIM mode. H. MAKING DILUTIONS: If the samples exceed the upper calibration range for the analysis, one of the following procedures may be used to estimate the high level concentrations. 1. Dilution Procedure A (dilution of the derivatized sample by reconstitution solvent): This option may be used only if the analytes in the sample were completely derivatized (see NOTE below). If derivatization was complete, transfer an aliquot of the sample from the GC vial (e.g. 0.2 mL for a 1:5 dilution) to a clean GC vial and dilute with reconstitution solvent (e.g. 0.8 mL for a 1:5 dilution), cap vial, mix, and reanalyze. However, dilution also dilutes the internal standard, and this procedure is useful only if the GC peak area for the internal standard is sufficiently measurable and the calibration curve is reasonably linear. Dilutions probably should not exceed a factor of 10. If this approach is used it is not necessary to enter a dilution factor in step 19 (V3/ V4) since both internal standard and analyte are diluted equally. The accuracy of this dilution procedure depends upon the linearity of the calibration curve in the extrapolated region beyond the upper end of the calibration curve. NOTE: Determination of Incomplete Derivatization: Incomplete derivatization can be caused by water, glycols, a large excess of analyte, or other contaminants that interfere with or compete for the derivatization reagent. If any one of the following symptoms appears, use Dilution Procedure B described below. a. An “oily” film (i.e., apparently viscous liquid) or unusual residue (e.g. grit) remains after being blown-down under nitrogen after derivatization (step 11d). This may be due to the presence of water, glycols, detergents, salts, or other contaminants. Incomplete derivatization has been observed with such residues. b. A very large (off scale) GC peak for any one of the derivatives (e.g. pseudoephedrine, a precursor for methamphetamine) indicates the possibility of incomplete derivatization for other analytes (e.g. methamphetamine) due to competition for the derivatization reagent. c. A smaller than usual GC peak area for the internal standard (<50% of the average) in undiluted samples suggests that something was competing for or inhibiting the derivatizing reagent. Such inhibition or competition for the internal standard will be experienced by the target analyte as well. d. Incomplete derivatization can be confirmed by the obvious presence of a GC peak for an underivatized target analyte. Underivatized analytes are not always detectable. Ephedrines usually do not show up on DB-5 capillary columns in this method, but GC peaks for underivatized secondary amines (e.g. methamphetamine) and for high levels of underivatized primary amines (e.g. amphetamine) can be detected, usually as irregularly shaped GC peaks, depending upon GC column conditions.

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition

Method rev. 1.1.1 METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 30 of 31

e. The problem of incomplete derivatization can be minimized by the use of an isotopic (e.g. deuterated) analog for each of the target analytes as the internal standard for that compound. This allows quantification in spite of incomplete derivatization. 2. Dilution Procedure B (dilution of a smaller aliquot of the original desorbate): If the sample was not completely derivatized or if large dilutions are needed (e.g. greater than about 1:5), the following procedure can be used. See NOTE in part 1 above. The procedure may also be used if derivatization was determined to be complete. a. Dilute an aliquot of the original aqueous acid desorbate of the wipe sample to 10 mL with desorption solution from a simulated sample blank, and re-extract. Add both the aliquot to be diluted and the diluting solution from the simulated blank directly to a clean 25-mL glass centrifuge tube (step 7f ) and proceed to step 8. For example, to make a 1:10 dilution, transfer 1 mL of original desorbate to the 25-mL tube and dilute with 9 mL from a simulated sample blank. b. The simulated sample blank should be prepared identically to the sample being diluted, using the same volumes of internal standard spiking solution and desorption solution that were used with the sample in the original desorption. For example, if the original sample was desorbed with 40 mL desorption solution with 80 μL of added internal standard spiking solution, then prepare the simulated blank in the same way. The volume of wetting alcohol is estimated (e.g. about 3 mL per 3”×3” (7.5 cm x 7.5 cm) 12-ply cotton gauze wipe). Include a dilution factor (V3/ V4) in the calculations in step 19 (e.g. V3/ V4 = 10 mL divided by the volume in mL of original desorbate diluted to 10 mL with solution from the simulated blank). The dilution factor in the above example is 10 mL/ 1 mL or 10. c. Correct for differences in internal standard spiking solution volumes in step 19 (if applicable) using for V1 the volume of internal standard spiking solution which was added to the original undiluted sample. NOTE: This dilution procedure gives quantitative results only if the residual volume of methanol (or isopropanol) used for wetting the sample wipes was exactly the same as the volume used in preparing the calibration standards (normally about 3 mL, see Table 7). Deviations of a few milliliters in residual wetting alcohol will not affect the results for undiluted samples, but will amount to an error of a few percent in the final results of samples that are diluted. d. The potential error due to differences in residual wetting solvent can be estimated for specific volumes of desorption solution and wetting alcohol. Assume the sample wipes and calibration standards are both desorbed in 30 mL of desorption solution and 3 mL of alcohol is added to the calibration standards. The potential error in volume (and final results) in the samples is approximately ±3 % (inversely proportional) per mL difference in the residual alcohol in the samples (i.e., ±1 mL difference in 33 mL). For 40 mL of desorption solution and 4 mL of alcohol added to the calibration standards, the error is ±2 % for every mL difference (i.e., ±1 mL difference in 44 mL). However, since the volume of residual wetting alcohol is not known and cannot be determined once the sample wipe has been desorbed, the actual error cannot be determined. However, the maximum possible error can be calculated. Since the maximum amount of alcohol that a recommended wipe can hold is about 6 mL when saturated (dripping wet), there can only be a deviation of plus or minus 3 mL from the 3 mL alcohol added to the calibration standards. Therefore, the maximum error in a result due to differences in the volume of residual alcohol in a cotton gauze sample compared to the standards can only be three times the error for a 1 mL difference in volume. Since the error for ±1 mL is ±3.03%, the maximum error for ±3 mL is three times larger, or ±9.1%. In practice, the error will be less than this because it is unlikely that the gauze samples will be completely dry or completely saturated after squeezing out the excess alcohol and wiping a surface. The practical amount of alcohol that remains in the wipes when the excess is squeezed out is between 1 and 2 mL. This translates into an error that is between +3% and +6% in the final results for diluted samples. Undiluted samples will not be affected. This error is within the overall accuracy for the method for methamphetamine.

Method rev. 1.1.1

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 31 of 31

3. Dilution Procedure C (dilution of desorbates from dried samples): Dilution errors for over-range samples may be corrected by knowing the exact amount of residual alcohol in the samples. The volume (or weight) of residual solvent in each gauze wipe might be determined by the difference between a wet weight and dry weight. Better yet, the error might be eliminated for diluted samples by adding, after the samples are dried (without taking any weight), the same known volume of wetting alcohol that is added to the calibration standards (i.e., 3 mL). Thereafter, if any samples need dilution, there will be no dilution errors due to differences in residual alcohol, because all samples and standards will have the same volume of alcohol and total volume of desorption solution. However, air drying of the samples is not recommended because of the possible loss of methamphetamine due to its volatility when it is not in the salt form, which form cannot be assured in field samples. Also, manipulating the samples for weighing and drying might introduce contamination. Drying is not recommended as a procedure for analytes having a vapor pressure high enough to be lost in the process, or that tend to form azeotropes with alcohols; this is especially important when the critical action levels for remedial cleanup are at the lower end of the method calibration range. Drying is not an option if the samples have already been desorbed.

NIOSH Manual of Analytical Methods (NMAM), Fifth Edition

Method rev. 1.1.1