D5449
The laboratory's allegation of compliance is not credible and evidence of correction is not acceptable.

Although the laboratory's submitted protocol indicates that an external positive QC material would be included in each run of patient Chlamydia trachomatis/Neisseria gonorrhoeae (CT/NG) testing, the laboratory provided no documentation of the positive QC material used, how the statistical parameters of the positive QC material would be determined, how the positive QC test result will be documented, and whether laboratory staff has been trained on this new protocol.

To ensure the deficient practice does not recur, the laboratory indicated that quarterly audits will be performed and suggested that the audits results would be reviewed within the laboratory's QMPI Program. However, the laboratory did not establish the procedure by which these quarterly audits are to be conducted. In its submission, the laboratory indicates that a "tracer audit may [emphasis added] be used," but did not provided a protocol for a "tracer audit," the means by which a "tracer audit" would be documented, and whether the results of a "tracer audit" would be the information reviewed by the QMPI Program.

The laboratory failed to address and provide acceptable evidence of correction consisting of: what measure has been put in place or what systemic changes have been made to ensure the deficient practice does not recur; and, how the corrective action(s) is being monitored to ensure the deficient practice does not recur.

D5469
The laboratory's allegation of compliance is not credible and evidence of correction is not acceptable.

Finding #1
Although the laboratory's submitted protocol indicates that the stated values of new commercially assayed CBC QC materials were to be verified through parallel testing against QC materials in use, the laboratory provided no documentation indicating that this protocol had been effectuated, no information as to how the results of the parallel testing will be documented, and no information as to whether laboratory staff has been trained on this new protocol.

To ensure the deficient practice does not recur, the laboratory indicated that quarterly audits will be performed and suggested that the audits results would be reviewed within the laboratory's QMPI Program. However, the laboratory did not establish the procedure by which these quarterly audits are to be conducted. In its submission, the laboratory indicates that a "tracer audit may [emphasis added] be used," but did not provided a protocol for a "tracer audit," the means by which a "tracer audit" would be documented, and whether the results of a "tracer audit" would be the information reviewed by the QMPI Program.

The laboratory failed to address and provide acceptable evidence of correction consisting of: what measure has been put in place or what systemic changes have been made to ensure the deficient practice does not recur; and, how the corrective action(s) is being monitored to ensure the deficient practice does not recur.