The laboratory failed to adequately address this deficiency and provide acceptable evidence of correction consisting of the required documentation and information set forth above and in our January 25, 2016 letter.

D5311
The laboratory's allegation of compliance is not credible and evidence of correction is not acceptable.

The submission references "Ex. A, Tab 4, § 7.1.1.1.d.2." We found no such reference contained in the materials provided to CMS.

The laboratory concluded that "this issue is not likely to affect patients" because its mislabel error rate is comparable to published industry mislabel error rates. No other information related to possible patient outcomes was provided. Even though the laboratory's error rate may be comparable, the laboratory must still pursue any corrective action(s) for patients that may have been affected by this deficient practice.

To ensure the deficient practice does not recur, the laboratory indicated that quarterly audits will be performed and suggested that the audits results would be reviewed within the laboratory's QMPI Program. However, the laboratory did not establish the procedure by which these quarterly audits are to be conducted. In its submission, the laboratory indicates that a "tracer audit may [emphasis added] be used," but did not provided a protocol for a "tracer audit," the means by which a "tracer audit" would be documented, and whether the results of a "tracer audit" would be the information reviewed by the QMPI Program.

The laboratory failed to adequately address this deficiency and provide acceptable evidence of correction consisting of the required documentation and information set forth above and in our January 25, 2016 letter.

D5391
The laboratory's allegation of compliance is not credible and evidence of correction is not acceptable.

Finding #1
The submission references "Ex. A, Tab 4, § 7.1.1.1.d.2." We found no such reference contained in the materials provided to CMS.

To ensure the deficient practice does not recur, the laboratory indicated that quarterly audits will be performed and suggested that the audits results would be reviewed within the laboratory's QMPI Program. However, the laboratory did not establish the procedure by which these quarterly audits are to be conducted. In its submission, the laboratory indicates that a "tracer audit may [emphasis added] be used," but did not provided a protocol for a "tracer audit," the means by which a "tracer audit" would be documented, and whether the results of a "tracer audit" would be the information reviewed by the QMPI Program.

6