PARTICULATES NOT OTHERWISE REGULATED, TOTAL
0501
DEFINITION: total aerosol mass
CAS: NONE
RTECS: NONE
METHOD: 0501, Issue 1
EVALUATION: FULL
Issue 1: 8 May 2015
OSHA:
15 mg/m³, total dust (inert or nuisance dust; particles not otherwise regulated) [1]
NIOSH:
no REL
For other OELs and guidelines: See references [2,3]
PROPERTIES:
Contains no asbestos and less than 1% quartz
SYNONYMS:
Nuisance dusts; particles not otherwise specified (PNOS)
SAMPLING
SAMPLER:
INTERNAL CAPSULE
(tared 37-mm, 2- to 5-µm PVC filter melded to PVC housing) in 37-mm 2-piece cassette
(tared 37-mm, 2- to 5-µm PVC filter melded to PVC housing) in 37-mm 2-piece cassette
FLOW RATE:
1 to 2 L/min
VOL-MIN:
17 L @ 15 mg/m³
-MAX:
333 L @ 15 mg/m³
SHIPMENT:
Routine
SAMPLE STABILITY:
28 days minimum
BLANKS:
Minimum of 2 field blanks per batch
ACCURACY
RANGE STUDIED:
0.1 to 4 mg per sample
BIAS:
0.058 [4]
OVERALL PRECISION (ŜrT):
0.059 [4]
ACCURACY:
± 15.5%
MEASUREMENT
TECHNIQUE:
GRAVIMETRIC (INTERNAL CAPSULE WEIGHT)
ANALYTE:
Airborne particulate material
BALANCE:
0.001 mg sensitivity; use same balance before and after sample collection
CALIBRATION:
National Institute of Standards & Technology Class S-1.1 weights or ASTM Class 1 weights
RANGE:
0.25 to 5 mg per sample
ESTIMATED LOD:
0.075 mg per sample
PRECISION(S1 ):
0.031 @ ≈2 mg per sample [4]
APPLICABILITY:
The working range is 2.5 to 50 mg/m³ for a 100-L air sample. This method is nonspecific and determines the ‘total’ dust concentration to which a worker is exposed.
INTERFERENCES:
Moisture and static electricity can affect gravimetric measurements. Humidity control and minimization of static effects are addressed in this procedure.
OTHER METHODS:
The method is similar to Method 5100 for carbon black [5]. This method is preferred over 0500, Issue 2 [6]. OSHA method PV2121 describes a similar procedure (but for respirable sampling) using an alternative sampler design [7].
NIOSH Manual of Analytical Methods (NMAM), Fifth Edition