ALLYL CHLORIDE 1000 CH2=CHCH2Cl   MW: 76.53   CAS: 107-05-1   RTECS: UC7350000 METHOD: 1000, Issue 2

EVALUATION: FULL

OSHA: 1 ppm NIOSH: 1 ppm; STEL 2 ppm ACGIH: 1 ppm; STEL 2 ppm

(1 ppm = 3.13 mg/m³ @ NTP)

Issue 1: 15 February 1984 Issue 2: 15 August 1994

PROPERTIES: liquid; d 0.938 g/mL @ 20 °C; BP 45 °C; MP –135 °C; VP 39.3 kPA (295 mm Hg) @ 25 °C; explosive range 3.3 to 11% v/v in air

SYNONYMS: 3-chloro-1-propene; 1-chloro-2-propene SAMPLING SAMPLER: SOLID SORBENT TUBE (coconut shell charcoal, 100 mg/50 mg)

MEASUREMENT TECHNIQUE:

GAS CHROMATOGRAPHY, FID

ANALYTE:

allyl chloride

DESORPTION:

1 mL benzene; stand 30 min

INJECTION VOLUME:

5 µL

FLOW RATE: 0.01 to 1 L/min VOL-MIN: -MAX:

16 L @ 1 ppm 100 L

SHIPMENT: routine SAMPLE STABILITY: at least 6 weeks @ 25 °C

TEMPERATURE-INJECTION: 185 to 200 °C -DETECTOR: 250 °C -COLUMN: 160 to 200 °C

BLANKS:

CARRIER GAS:

N2, 30 mL/min

COLUMN:

stainless steel, 1.2 m × 6-mm OD; 50/80 mesh Porapak Q or equivalent

CALIBRATION:

standard solutions of allyl chloride in benzene

RANGE:

0.05 to 1.5 mg per sample

2 to 10 field blanks per set ACCURACY

RANGE STUDIED: BIAS:

1.8 to 7.2 mg/m³ [1] (100-L samples) 6.4%

OVERALL PRECISION ( ): 0.071 [1]

ESTIMATED LOD: 0.01 mg per sample ACCURACY:

±16.6%

PRECISION ( ):

0.023 [1]

APPLICABILITY: The working range is 0.16 to 3 ppm (0.5 to 10 mg/m³) for a 100-L air sample. The method is applicable to short-term samples taken at 1 L/min. The upper limit of loading depends on the concentrations of allyl chloride and other substances in the air, including water vapor. INTERFERENCES: None identified. OTHER METHODS: This is Method S116 in a revised format [2].

NIOSH Manual of Analytical Methods (NMAM), Fourth Edition