PENTACHLOROETHANE
2517
CHCl2CCl{{{1}}}
MW: 202.30
METHOD: 2517, Issue 2
EVALUATION: FULL
Issue 1: 15 May 1985
Issue 2: 15 August 1994
Issue 2: 15 August 1994
OSHA:
no standard
NIOSH:
handle with caution
ACGIH:
no standard (1 ppm = 8.27 mg/m³ @ NTP)
PROPERTIES:
liquid; d 1.671 g/mL @ 25 °C; BP 162 °C; VP 450 Pa (3.4 mm Hg; 4500 ppm) @ 20 °C
SYNONYMS:
Pentalin; ethane pentachloride.
SAMPLING
SAMPLER:
SOLID SORBENT TUBE
(Porapak R, 70 mg/35 mg)
(Porapak R, 70 mg/35 mg)
FLOW RATE:
0.01 to 0.2 L/min
VOL-MIN:
1 L @ 0.1 mg/m³
10:
10 L
SHIPMENT:
routine
SAMPLE STABILITY:
stable ≥ 7 days @ 25 °C;
≥ 21 days @ 0 °C in the dark
BLANKS:
2 to 10 field blanks per set
ACCURACY
RANGE STUDIED:
0.014 to 5.37 mg/m³[1]
(3-L samples)
(3-L samples)
BIAS:
not significant [1]
OVERALL PRECISION ([1]):
0.075 [1]
ACCURACY:
± 15%
MEASUREMENT
MEASUREMENT:
GAS CHROMATOGRAPHY,63Ni ECD
ANALYTE:
pentachloroethane
DESORPTION:
2 mL hexane; 30 min ultrasonic
TEMPERATURE-INJECTION:
240 °C
-DETECTOR:
250 °C
-COLUMN:
70 °C
INJECTION VOLUME:
5 µL
CARRIER GAS:
5% methane, 95% argon; 30 mL/min
DETECTOR PURGE GAS:
5% methane, 95% argon if required; 80 mL/min
COLUMN:
nickel, 2-mm ID x 2 m, packed with 3% OV-17 on 100/120 mesh Chromosorb WHP
CALIBRATION:
standard solutions of pentachloroethane in hexane
RANGE:
0.04 to 40 µg per sample
ESTIMATED LOD:
0.004 µg per sample [2]
PRECISION ([2]):
0.019 [1]
APPLICABILITY: The working range is 0.01 to 20 mg/m³ (0.001 to 2.4 ppm) for a 5-L air sample.
INTERFERENCES: 1,2,2,3-Tetrachloropropane interferes with the analyte GC peak but is not likely to be present in most samples. Large concentrations of 1,1,2,2-tetrachloroethane may obscure the analyte GC peak.
OTHER METHODS: This revises P&CAM 335 [2].
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition, 8/15/94