The submission also states that "the lab has conducted training on these procedures to ensure that the practice is consistent with them," but the laboratory did not include complete training documentation to support this assertion.

Because the laboratory has not shown whether it can follow its own validation protocols, it was uncertain whether the laboratory's quality assessment mechanisms could monitor the laboratory's corrective actions and ensure this deficient practice does not recur.

The laboratory failed to adequately address this deficiency and provide acceptable evidence of correction consisting of the required documentation and information set forth above and in our January 25, 2016 letter.

D5423
The laboratory's allegation of compliance is not credible and evidence of correction is not acceptable.

Although the laboratory's submitted protocol (Ex. A, Tab 9) included a method verification procedure which addressed verification of accuracy, precision, and reportable range, the laboratory provided no documentation showing that this protocol had been followed to re-verify performance specifications. The submission also states: "Before the lab resumes any test on the Advia XPT, the lab will ensure that the test has been re-verified pursuant to the lab's improved method verification procedures that have been approved by the laboratory director." The completion date for the correction of this deficiency was indicated on the submitted Allegation of Compliance as February 12, 2016 which, in order to determine correction, should include the re-verification documentation for the Advia XPT. However, the laboratory provided no documentation that the re-verification had been performed.

We note that in the laboratory's submission related to test performance specifications for the Advia XPT instrument, it was the laboratory's policy to accept %CV values that were up to 1.5 times the %CV values stated by the instrument manufacturer for any given chemistry analyte. In the submission, we found no explanation as to why 1.5 times the manufacturer's %CV was acceptable to the laboratory. We also found no explanation as how "normal patient distribution" relates to determining if patients were affected or potentially affected.

The laboratory's procedure as well as the method verification for alkaline phosphatase was reviewed at the onsite survey and the reportable range was documented as 0 - 1100 IU/L thus modifying the test. At the time of survey, the laboratory stated that the reportable range was 10 1100 IU/L per the manufacturer, and that the reportable range of 0 - 1100 IU/L was an error. The laboratory confirmed that the lowest reportable value established by the manufacturer was 10 IU/L. The Patient Impact Analysis submitted by the laboratory at Ex. B, Tab 6 now indicates that the reportable range should be 5 - 1100 IU/L which is still lower than the range established by the manufacturer; however, the laboratory did not submit any documentation to support this new reportable range.

The Patient Impact Analysis also indicates that review of the PT showed a negative bias for PT samples CAP C-A 2015 and CAP-B 2015, but the laboratory determined that no corrected reports