the summary or raw data, so we were unable to determine when the correlation studies occurred. The submission for SHBG also included information on one document with box plots that included three devices and another document that included box plots of six devices, so we were unable to determine what devices were actually used for SHBG testing.

The laboratory's submission offered no response to correlation studies for Vitamin B12 and no explanation for the disparity between the devices used for QC and the devices used for patient testing. We also noted that there was no explanation as to where the total allowable error (TAE) of 24% on the raw data documentation came from or if any TAE was calculated for the submitted data. The laboratory used the median result for each proprietary device to determine if TAE was acceptable, but did not provide an explanation why the median results were used.

To ensure the deficient practice does not recur, the laboratory indicated that quarterly audits will be performed and suggested that the audits results would be reviewed within the laboratory's QMPI Program. However, the laboratory did not establish the procedure by which these quarterly audits are to be conducted. In its submission, the laboratory indicates that a "tracer audit may [emphasis added] be used," but did not provided a protocol for a "tracer audit," the means by which a "tracer audit" would be documented, and whether the results of a "tracer audit" would be the information reviewed by the QMPI Program.

The laboratory failed to address and provide acceptable evidence of correction consisting of: what measure has been put in place or what systemic changes have been made to ensure the deficient practice does not recur; and, how the corrective action(s) is being monitored to ensure the deficient practice does not recur.

D5779
The laboratory's allegation of compliance is not credible and evidence of correction is not acceptable.

In Ex. F, the laboratory provided lists of patient specimen accession numbers for which the laboratory intended to issue corrected test reports. The laboratory provided no documentation to indicate corrected reports were issued.

To ensure the deficient practice does not recur, the laboratory indicated that quarterly audits will be performed and suggested that the audits results would be reviewed within the laboratory's QMPI Program. However, the laboratory did not establish the procedure by which these quarterly audits are to be conducted. In its submission, the laboratory indicates that a "tracer audit may [emphasis added] be used," but did not provided a protocol for a "tracer audit," the means by which a "tracer audit" would be documented, and whether the results of a "tracer audit" would be the information reviewed by the QMPI Program.

The laboratory failed to adequately address this deficiency and provide acceptable evidence of correction consisting of the required documentation and information set forth above and in our January 25, 2016 letter.

D5787