The laboratory's allegation of compliance is not credible and evidence of correction is not acceptable.

The laboratory submitted no written procedure for the documentation of the lot of bacteriology media used to test any given patient specimen, and no information as to whether laboratory staff has been trained on this new protocol.

In its submission, the laboratory references Ex. L, Tab 6 which states: "When [media quality control] testing is completed the results are logged into an exel sheet with lot number and expiration date for each patient (see attached report)." We find no "attached report" in the submission and, therefore, no documentation indicating the laboratory's corrective action has been effectuated.

To ensure the deficient practice does not recur, the laboratory indicated that quarterly audits will be performed and suggested that the audits results would be reviewed within the laboratory's QMPI Program. However, the laboratory did not establish the procedure by which these quarterly audits are to be conducted. In its submission, the laboratory indicates that a "tracer audit may [emphasis added] be used," but did not provided a protocol for a "tracer audit," the means by which a "tracer audit" would be documented, and whether the results of a "tracer audit" would be the information reviewed by the QMPI Program.

The laboratory failed to address and provide acceptable evidence of correction consisting of: what measure has been put in place or what systemic changes have been made to ensure the deficient practice does not recur; and, how the corrective action(s) is being monitored to ensure the deficient practice does not recur.

D5791
The laboratory's allegation of compliance is not credible and evidence of correction is not acceptable.

Finding #1
The laboratory submitted a protocol for the SensoScientific Monitoring in Ex. A, Tab 24. The response does not specifically address the days that the freezers did not meet acceptable temperatures as documented in the SensoScientific Monitoring Audit Node Log. We could not determine if the new procedure has been effectuated as no documentation was submitted to show that the unacceptable freezer temperatures were addressed. In addition, the laboratory states that training has occurred, but it was not clear who should be trained as "laboratory supervisor or designee" was not defined. Training records in Ex. A, Tab 30, included only training documents for one general supervisor. Several freezers were identified as "not used in patient testing" (see Ex. N, Tab A); however, based on interviews during the onsite visit, these freezers were identified as being used for CLIA activities. The submission does not include a response or data to support the laboratory's assertion that certain freezers are "not used in patient testing."

The laboratory submitted an article for the use of Mean Kinetic Temperatures (MKT) (Ex. N, Tab 1). This article relates to the handling, storage, and distribution of temperature sensitive pharmaceuticals, not clinical laboratory specimens, reagents, reference materials, etc. We are