The laboratory's allegation of compliance is not credible and evidence of correction is not acceptable.

The laboratory submitted protocols relating to testing personnel training (Ex. A, Tab 16, §§ 4.5.3 and 6.4) which state the following:

  • § 4.5.3 "Testing personnel/trainee are responsible to ensure that training is properly documented using CL-FRM-03016-F4, Training Attendance Form, approved by the designated trainer and stored in the relevant binder."
  • § 6.4 " . . .It is the responsibility of the trainee and trainer to ensure the SOP is accurate."

However, the regulations require that the laboratory director, not the testing personnel, is responsible for ensuring that prior to testing patient specimens, all personnel have the appropriate training and have demonstrated that they can perform all testing operations to provide and report accurate results.

Training documentation submitted in Ex. A, Tab 11 only includes training on protocols related to testing personnel qualification requirements and delegation of responsibilities, as well as job descriptions. The laboratory provided no documentation or plan to train or retrain testing personnel on the Theranos Proprietary Device. Training documents submitted in Ex. A, Tab 17 only addressed review of the document control and training and competency protocols, but did not include training or retraining documentation for personnel performing testing in the venipuncture lab nor did it include an investigation if patient were affected or potentially affected when untrained or partially trained testing personnel were performing patent testing.

To ensure the deficient practice does not recur, the laboratory indicated that quarterly audits will be performed and suggested that the audits results would be reviewed within the laboratory's QMPI Program. However, the laboratory did not establish the procedure by which these quarterly audits are to be conducted. In its submission, the laboratory indicates that a "tracer audit may [emphasis added] be used," but did not provided a protocol for a "tracer audit," the means by which a "tracer audit" would be documented, and whether the results of a "tracer audit" would be the information reviewed by the QMPI Program.

The laboratory failed to adequately address this deficiency and provide acceptable evidence of correction consisting of the required documentation and information set forth above and in our January 25, 2016 letter.

D6108
The laboratory's allegation of compliance is not credible and evidence of correction is not acceptable.

See our review of 06115.

D6111
Based on the laboratory's submission, this requirement is met.