This page has been proofread, but needs to be validated.

D6115
The laboratory's allegation of compliance is not credible and evidence of correction is not acceptable.

The laboratory submitted a protocol for a Master Validation Plan at Ex. E, Tab 30B which was a different protocol than presented to the surveyor. In Ex. E, Tab 1, under the Accuracy section, the submission indicates that "some of the immunoassays on the TSP [Theranos Proprietary System] were 'calibrated' or 'corrected' to remove systemic bias between the two methods." It is not clear what specific biases were used for each analyte which used a corrected results or if the corrected or uncorrected results was reported. In addition, the Accuracy section described that the "samples spanned the medical decision levels (MDL)". We are unclear as to the laboratory's definition of "medical decision limits." No definition of this phrase was found in Section 3—Definitions of the "Method Validation" protocol or in Ex. E. We also note that the submitted protocol does not include a reference to "medical decision limits" as a criterion for determining accuracy.

We also note that the laboratory's submission provided no indication why the validation report for SHBG had an effective date of 7/14/14, but was not approved by the laboratory director until 9/19/15, even though patient testing began 7/28/14.

The laboratory's submission in Ex. E, Tab 1 states:

The procedure and acceptance criteria outlined in the "Guidance for Industry: Bioanalytical Method Validation, 2001" document from FDA was followed to establish the analytical measurement range (AMR) (reportable range) for these immunoassays run on the TPS. Theranos did not follow the validation plan for immunoassays (ELISAs) as outlined in Master Validation Plan for ELISA Assays on Theranos Devices.

The laboratory's response does not include a copy of the FDA protocol, and does not state in the submitted master validation protocol, CL PLN-14002, Rev. A, that the FDA protocol should be used to determine AMR.

The laboratory submitted a protocol for a Master Validation Plan at Ex. E, Tab 30B which requires the %CV of replicates to be not more than 20% (25% at the lower and upper limits of detection). The submitted protocol was different than the protocol presented to the surveyor at the time of the onsite survey which required the %CV of replicates to be not more than 15% (20% at the lower and upper limits of detection). The %CV in the protocol presented to the surveyor at the time of the onsite survey matched the %CV of the predicate devices. Based on the protocol submitted in the submission, it is unclear as to why the TPS, which is covered by the Master Validation Plan, can now have %CV's for replicates greater than the predicate devices. No explanation for the change in %CV was submitted. In addition, we noted that the information submitted in Ex. E, Tab 1 did not include a 20-day precision study as required by the submitted protocol.

The laboratory's submitted protocol requires that 120 samples be used to determine Reference Intervals (see Ex. E, Tab 30B), Ex. E, Tab 1 states:

38