However, the validation plan noted that the reference range would be established with 120 samples per partition. However, the study followed CLSI procedure for transferring a reference range—the process by which one adapts a previously established reference range to a new analytical method. Furthermore, the transferred reference range was verified following CLSI guidelines—namely "the process by which one ensures, with reasonable confidence, using a relative small number of reference individuals (eg, n=20), that a reference interval established elsewhere, or transferred from another study, can be used locally.

We question the laboratory's procedure for "transferring a reference range" since the laboratory provided no reference document to support "transferring a reference range." We note that the submitted protocol does not state in Section 18, Reference Range, that the laboratory can "transfer" a reference range. It is unclear as to how the laboratory determined that the predicate method's reference range could be adapted or "transferred" to the Theranos Proprietary Devices, which used different methodology than the predicate method.

The laboratory's submission does not indicate that the % Recovery and Allowable Bias citations had been addressed.

Ex. E contains a list of patient specimen accession numbers for which the laboratory "identified reports", but did not specify, or submit documentation to show that corrected reports were generated and issued.

Because the laboratory has not shown that it can follow its own validation protocols, it is unclear whether the laboratory's quality assessment mechanisms could monitor the laboratory's corrective actions and ensure this deficient practice does not recur.

The laboratory failed to adequately address this deficiency and provide acceptable evidence of correction consisting of the required documentation and information set forth above and in our January 25, 2016 letter.

D6124
The laboratory's allegation of compliance is not credible and evidence of correction is not acceptable.

Ex. A, Tab 16, § 7.3 of the submission contains a competency assessment protocol, Annual Competency. However, the protocol does not address the cited deficiency that direct observation was required for instrument maintenance and function checks. The protocol at § 1.1.7 (page 12) allows for a choice of validation method of which direct observation is one choice. The submitted laboratory protocol also requires the use of CL FRM 03016-F3 for annual competency. This form also allows the assessor to choose the validation method for competency assessment. Neither the submitted protocol nor the form to document competency assessment indicate that the six required procedures for CLIA competency assessment are required.

Ex. L, Tab 4 of the submission states: "All testing personnel currently performing tests have completed competency testing with direct observation of, among other things, maintenance and